32
WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
7.2.5
Set interval mode
When operating the
AlphaGUARD
in the interval mode, the measurement takes
place just in defined time period for a defined sample time, e.g. every 60 minutes
for 30 minutes. With this mode an extended battery life time can be achieved for
flow operation.
The interval mode can be activated by choosing “ON” pressing the
Scroll up
or
Scroll down
button and pushing the
Enter
button for 2 sec for acknowledgement.
The interval mode can be configured by the following parameters:
Pre-sample
time:
Activated pump to flush the measurement chamber before
sampling is started
During the Pre-Sample time no measurement is performed
Sample time:
Activated pump and running measurement
Time period:
The time period includes the complete time interval (from the
start of the sampling time to the start of the next sampling time)
Start time:
Start time of the first sample time
Set Interval mode 10min
1 1 . 0 8 . 1 6 1 2 : 2 7 : 4 7 4 m i n
P r e - s a m p l e t i m e :
M e a s u r i n g t i m e :
3 0 m i n
I n t e r v a l t i m e :
1 2 0 m i n
I n t e r v a l s t a r t :
1 6 : 0 0
I n t e r v a l m o d e :
1 0 m i n
O n
Figure 23: Settings for interval mode
-
The interval time has to be longer than the pre-sample time + the measuring
time.
-
Within the sample time the measurement values are stored according to
the selected operation mode. In the meantime, no measurement values
are stored and the read-out data series shows gaps for these periods.
Note
Содержание ALPHAGUARD D2000
Страница 12: ...12 WWW BERTIN INSTRUMENTS COM USER MANUAL Figure 3 AlphaGUARD front side Figure 4 AlphaGUARD handle side...
Страница 82: ...82 WWW BERTIN INSTRUMENTS COM USER MANUAL Figure 65 Range of delivery 3 1 10 11 5 4 7 2 6 8 9 12 13...
Страница 93: ...93 WWW BERTIN INSTRUMENTS COM USER MANUAL Flow diagrams of the AlphaGUARD menu 14 3 1 Acquisition windows...
Страница 113: ...113 WWW BERTIN INSTRUMENTS COM USER MANUAL...