111
WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
0x00002000
Humidity sensor init
failed
Humidity/Temperature sensor
defective, please contact your local
Bertin dealer, press enter to continue
0x00004000
Clock battery low
0x00010000
Internal pump module
failed to initialize
Pump module defective, please
contact your local Bertin dealer, press
enter to continue
0x00020000
Failed to initialize GM
board
Dose rate module not found or
defective, please contact your local
Bertin dealer, press enter to continue
0x00040000
Failed to initialize
bluetooth module
Volatile Errors, device still operational
0x01000000
Battery voltage is
below operation level
Battery low, connect charger to
operate the device, device will switch
off in 9 seconds
0x02000000
Clock needs to be set
System clock needs to be set, press
<enter> to continue
0x04000000
File system not present
Filesystem corrupted, erase memory
to repair filesystem, press <enter> to
continue
0x08000000
Database consistency
test failed
Recommendation to read out device
and erase memory
0x10000000
High voltage
generation failed