21
WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
With the flow-through adapter (screw cap) the
AlphaGUARD DF2000
can be
operated in flow mode (open and in closed cycle).
Converting
AlphaGUARD DF2000
again for diffusion mode requires the above
steps in reverse order. It is sufficient to screw the diffusion adapter slightly by hand
as it has no sealing function.
To avoid contamination of the plain paper filter at the chamber entrance, an
external radon progeny filter (accessory) has to be connected to the tube in front
of the air inlet.
Figure 9: Using external radon progeny filter in front of the air inlet (flow mode)
Air
Operate the active adapter only with connecting in series
the external radon progeny filter (enclosed accessory);
otherwise there is acute danger of contamination of the
plain paper filter at the chamber entrance.
Important note
Содержание ALPHAGUARD D2000
Страница 12: ...12 WWW BERTIN INSTRUMENTS COM USER MANUAL Figure 3 AlphaGUARD front side Figure 4 AlphaGUARD handle side...
Страница 82: ...82 WWW BERTIN INSTRUMENTS COM USER MANUAL Figure 65 Range of delivery 3 1 10 11 5 4 7 2 6 8 9 12 13...
Страница 93: ...93 WWW BERTIN INSTRUMENTS COM USER MANUAL Flow diagrams of the AlphaGUARD menu 14 3 1 Acquisition windows...
Страница 113: ...113 WWW BERTIN INSTRUMENTS COM USER MANUAL...