11.4 Evaluation Unit LB 442
96
HV Settings
Set detector HV
With the standard setting ”0” the detector works with automatic
high voltage control. For service purposes, you can enter a high
voltage value between 500 V and 1500 V. In this case, the high
voltage control is disabled. The voltage control applied at the pho-
tomultiplier then corresponds to the entered value.
For the regular measurement mode you have to enter 0.
Save default
Here you can enter a high voltage start value. This allows rapid
control of the HV to the optimum operation point after power out-
age. The value can lie between 500 V and 1500 V.
Read default
The start value of the high voltage stored in the detector is dis-
played here.
Status request
o
Temp.PT100
The PT 100 temperature is displayed
.
o
NTC-Temp.
The temperature in the detector is displayed.
o
Pulses
The current count rate of the detector is displayed.
o
HV
The currently set high voltage is displayed
Reset Detector
A reset on the detector CPU can be carried out.
Pt 100 Adjust
With a 100 Ohm resistance at terminals 3/4 of the detector, the
PT100 input can be adjusted with the button combination "
sk2 &
clear
".
Содержание LB 442
Страница 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Страница 2: ......
Страница 117: ...Section 13 Appendix 115 Section 13 Appendix 13 1 Certificates 13 1 1 ATEX Certificate for Evaluation Unit...
Страница 118: ...13 1 Certificates 116...
Страница 119: ...Section 13 Appendix 117...
Страница 120: ...13 1 Certificates 118...
Страница 121: ...Section 13 Appendix 119...
Страница 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Страница 123: ...Section 13 Appendix 121...
Страница 124: ...13 1 Certificates 122...
Страница 125: ...Section 13 Appendix 123...
Страница 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Страница 127: ...Section 13 Appendix 125...
Страница 128: ...13 1 Certificates 126...
Страница 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Страница 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Страница 133: ...Section 13 Appendix 131...
Страница 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Страница 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Страница 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Страница 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...
Страница 148: ...13 5 Wiring Diagram 146 13 5 3 19 Frame...