6.2 Software Functions and System Configuration
54
Max. Basis Weight
For non-linear calibration curves you have to enter the maximum
weight per area and you have to select the function cubic for calcu-
lation. These square or cubic characteristic curves require absorp-
tion coefficients a2 and a3 in addition to the absorption coefficients
a1. This kind of calibration curve is required only in rare cases. In
such cases you then have to enter the maximum weight per area
at maximum load in g/cm
2
prior to calculation. Without this infor-
mation, you will get the following error message when you try to
perform a cubic calculation: "Enter weight per area".
The weight per area is calculated as follows:
M
(t/h)
F (g/cm
2
) = -------------------------------
v (m/s) x b (m) x 36
Example:
Measuring range:
0 – 200 t/h
Belt speed v:
1 m/s
Belt width b:
1 m
F = 5,55 g/cm²
Belt width
(0 to 9,99 m)
Enter the belt width in meters for calculation of the weight per a-
rea.
Product select [external/internal
Specify if the product is to be selected via the evaluation unit (=
internal
) or via digital inputs (=
external
, 18a and 18c, 20a and
20c).
Low load suppression
The low load entered here is suppressed and as long as the en-
tered value in t/h (kg/h) is not exceeded, the display shows
0 t/h
(kg/h)
.
Erase mass integ.
Yes
means: reset the mass integrator to zero for each new run.
No
means: The mass integrator continues to count from the value
reached last in each run
.
Содержание LB 442
Страница 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Страница 2: ......
Страница 117: ...Section 13 Appendix 115 Section 13 Appendix 13 1 Certificates 13 1 1 ATEX Certificate for Evaluation Unit...
Страница 118: ...13 1 Certificates 116...
Страница 119: ...Section 13 Appendix 117...
Страница 120: ...13 1 Certificates 118...
Страница 121: ...Section 13 Appendix 119...
Страница 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Страница 123: ...Section 13 Appendix 121...
Страница 124: ...13 1 Certificates 122...
Страница 125: ...Section 13 Appendix 123...
Страница 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Страница 127: ...Section 13 Appendix 125...
Страница 128: ...13 1 Certificates 126...
Страница 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Страница 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Страница 133: ...Section 13 Appendix 131...
Страница 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Страница 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Страница 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Страница 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...
Страница 148: ...13 5 Wiring Diagram 146 13 5 3 19 Frame...