![Berthold Crocodile miniWorkstation Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/berthold/crocodile-miniworkstation/crocodile-miniworkstation_operating-manual_2739253021.webp)
LB 925 Crocodile miniWorkstation
2. Important information
84024BA2 Rev. 03, 04/2020
Page 21 of 68
2.4
Further instructions
The information in this manual complies with the actual state of
knowledge at the publication date. When LB 925 is operated in
compliance with the instructions in this manual, there are no known
risks for the user, the environment or the quality of the
measurement results. However, the user should be aware of
situations that could result in serious damage. Always read the safety
instructions and precautions carefully!
CAUTION
The user must assure that the whole system of LB 925, software and
assay is subject to a continuous risk analysis done by the user.
Storage conditions
Before delivery or if the instrument is not used for a longer period of
time, store it in the original cardboard box in a dry dust-free
environment and protected from direct sunlight and significant
temperature fluctuations!
Storage temperature:
5-40°C up to 75% humidity (at 30°C), non-condensing
Transport conditions
-25° to +60°C, up to 75% humidity, non-condensing. Transport in
original cardboard box and free of liquids.
Quality control
It is considered good laboratory practice to run known internal
quality standards or laboratory samples as well as reagent kit
controls, attendant to the measurement runs, according to
instructions and specific recommendations included in the package
insert of the reagent kit or the standard laboratory protocol for the
test to be conducted. Samples should be obtained, treated and
stored following the instructions and recommendations of the kit
insert.
To ensure proper operation of the system, it is recommended to
operate Qualification Procedures at defined time intervals and to use
high quality equipment.
Berthold Technologies offers Validation Tools for the Crocodile
miniWorkstation, including the IQ/OQ/PQ Qualification Package and
the Absorbance TestPlate.