2. Important information
LB 925 Crocodile miniWorkstation
Page 14 of 68
84024BA2 Rev. 03, 04/2020
CAUTION: The liquids and test components
processed with the device may pose a hazard, e.g. a potential
biohazard. For this reason, the reagent inlets and the waste
outlet on the side are marked with these combined labels.
If liquid may have got inside the device, pull out the mains plug
immediately. Never operate the instrument if internal
components may have been in contact with liquids, as this may
result in electric shock or fire. Have the unit cleaned by an
authorized service center.
Prior to repair or maintenance, the instrument must be cleaned
and decontaminated to prevent service personnel from coming
into contact with potentially biohazardous materials or other
hazards.
All service and maintenance work beyond the scope of this
manual may only be carried out by service technicians
authorized by Berthold Technologies. Before further use, the
instrument must be reassembled and inspected in accordance
with the instructions in the service manual.
Only parts specified by Berthold Technologies may be used for
maintenance work.
Switch off the instrument before disconnecting the plug.
Always disconnect the plug before opening the device for
maintenance work or modifications.