LB 379
9
1.3
Device Configuration (Operating Modes)
The evaluation unit can be used for different tasks.
Density measurement without temperature compensation (TC)
Density measurement with temperature compensation
via Pt100 or
density measurement with temperature compensation via current input
Suspension measurement with any carrier liquid
solid density and liquid density are known
with or without TC
Suspension measurement with water as the carrier liquid
solid density is known
with water temperature compensation, the density of the water and
the concentration of the solid material are taken into account for temperature
compensation
Mass flow measurement without TC (in connection with a volume flow meter)
Mass flow measurement with TC via Pt 100 (in connection with a volume flow
meter)
Measurement modes
Continuous measurement
Discontinuous measurement
Batch operation via keyboard (press <run> button)
Batch operation via digital input
1.3.1
Setting the Detector Code
Internal device parameters are adjusted to suit the used scintillator size by setting
the detector code. The correct detector code is set at the factory and a change is
not normally required. A table with the detector codes to be used is found below:
Measuring Path
Detector Code
LB 379
2
LB 3602/3603
manual
For the measuring path LB 3602/3603 you must set the HV working point to manually
fixed value. If the value is not known you must evaluate the value by a plateau measure-
ment. The plateau measurement is descriped in the manual LB 474 respectively LB 444.
The HV working point must be entered in the evaluation unit at the mid of the plateau
curve.
1.3.2
Measuring Path Dimension
Input of the measuring path dimension in the evaluation unit:
Measuring
Path
without
Liner
with
PTFE - Liner
With
Rubber - Liner
LB 379
6.8
6.0
5.8
LB 3602/3603 5.1
-
-
Dimensions in cm
Содержание LB 3602
Страница 1: ...Measuring Path LB 379 and LB 3602 3603 Operating Manual 40437BA2 Rev No 00 12 2016...
Страница 2: ...LB 379 2...
Страница 4: ...LB 379 4...
Страница 31: ...LB 379 31 9 Dimensional Drawings 9 1Measuring Path LB 379 with DIN Flange...
Страница 32: ...LB 379 32 9 2Measuring Path LB 379 with ASA Flange...
Страница 33: ...LB 379 33 9 3Measuring Path LB 379 with Threaded Connector...
Страница 34: ...LB 379 34 9 4Measuring Path LB 3602 03...
Страница 35: ...LB 379 35 10 ATEX Certificate...
Страница 36: ...LB 379 36...
Страница 37: ...LB 379 37...
Страница 38: ...LB 379 38...
Страница 39: ...LB 379 39...
Страница 40: ...LB 379 40...
Страница 41: ...LB 379 41...
Страница 43: ...LB 379 BERTHOLD TECHNOLOGIES GmbH Co KG Calmbacher Str 22 D 75323 Bad Wildbad Germany www Berthold com...