7. Processing the Results
43
Format of Data Transfer
a) Data transfer to PC
Data is transferred in the following order:
Date
Time
Program
Limit pass
Limit fail
Measuring time
Location identification number
Test identification number
Result
(= integrated RLUs)
or ”Overload”
<CR>
(= carriage return)
<LF>
(= line feed)
If an overload occurs during a measurement, “
Overload
” is
transferred instead of a result (RLU value).
Each value is separated by using <TAB>.
Example:
03/23/98<TAB>08:32:58<TAB>ATP<TAB>10000<TAB>5000
<TAB>10<TAB>1<TAB>3<TAB>1556<TAB><CR><LF>
b) On-line data transfer to PC
Here, data are transferred in the same format and sequence
as in a) however, additionally 100 single RLU values are
transferred, which are positioned in front of the integrated
RLU value. Also the single RLU values are separated by
using <TAB>.
If the measurement is aborted, the values are transferred up
until stopping the measurement and ”
CANCEL
” will be
transferred.