5. Setting the Parameter
junior
30
o
Press the <
ENT
> key in order to activate the Input mode
and enter the RLU value for the [
limit fail
] which will
serve as the upper/lower threshold value to classify
results as
fail
.
o
Confirm the entry with the <
ENT
> key.
o
Move the cursor to the [
O.K.
] parameter and press the
<
ENT
> key.
o
Enter the RLU value for the [
limit pass
] which will serve
as the lower/upper threshold of the range that will be
used to classify results as
pass
(Acceptance range).
o
Be aware
that results that fall between the two
thresholds will be classified as
warn
(= Warning range).
o
Confirm the entry with the <
ENT
> key.
o
Move the cursor to the [
Measuring Time
] parameter
and press the <
ENT
> key.
o
Enter the desired measuring time in seconds and
confirm the entry with <
ENT
>.
o
When all individual parameters have been defined and
confirmed with the <
ENT
> key, press the <
ESC
> key in
order to return to the [
Settings
] sub-menu.
o
If you wish, you can select a different program (2-6) and
define its measurement parameters. Proceed as
described for [
ATP
].