6.
Measuring
junior
38
6.2
Measurement Procedure
o
Prepare the sample according
to the reagent manufacturer’s
instructions.
o
Select
the
desired
measurement protocol and if
necessary, enter a location, as
well as a test identification
number.
o
Open
the
cover
of
the
measuring chamber and insert
the test tube into the sample
holder.
o
Close the measuring chamber.
The cap must click into place.
The measuring chamber will
then be light tight and the
measurement can be started.
o
At the [
Measurement Start
]
menu option, press the <
ENT
>
key.
o
The measurement will starts.
The remaining measuring time
in seconds is displayed while
the measurement is being
taken.
o
On completion of the measurement the RLU value is
displayed. The measurement will also be classified
according to the threshold value settings (
pass
= O.K.,
fail
= Error range or
warn
= Warning range) and
indicated by the illumination of the appropriate LED.
Assay: ATP
Location: 1
Test: 1
Time: 6,5,4...
Assay: ATP
Location: 1
Test: 1
Result: pass
RLU: 111
Measurement Start
Assay: ATP
Location: 1
Test: 1
insert sample
ENT
ENT
Figure 6.2:
Measurement procedure