Figure 25
–
CDP Base at maximum vertical stance
9.
Positioning in desired installation area
a.
Position CDP attached to dollies in the desired area
b.
Lower system using your drill until it is about 1/3 of the way from the bottom of its
range of motion
c.
Remove stabilizer bar and move it to the top of the dollies
d.
Lower system so that the base is just above the floor and still supported by dollies
e.
While the base is horizontal and connected to dollies adjust the base into final position
i.
4.5” minimum distance from rear of CDP to closest wall or obstruction
ii.
12” minimum distance from CDP sides
to closest wall or obstructions
f.
Lower dollies completely until base is firmly resting on the ground.