1015002x00
105
Configuring and Operating the Driving Unit
en
8.6.5
Optimizing the Filling and Emptying of the Blood Pumps
8.6.5.1 Blood Pump Phases
8.6.5.2 Basic Procedure
Individual settings:
The parameter values must be determined and set individually
for each patient.
Preconditions:
Before setting the parameters, the status of the patient must be as-
certained (blood volume, MAP,PAP, CVP). Ensure that the patient has sufficient blood
volume. Patient blood pressure will need to be between 60 and 80 mmHg, but certain-
ly no higher than 80 mmHg. At higher pressures, the blood may not be completely driv-
en out of the blood pump. If necessary, reduce the patient’s blood pressure using
medication.
Goal:
Generate a blood flow that corresponds to the patient’s needs.
Choose settings in this order:
1. Rate - 2. Filling/Emptying (++ + - --) - 3. Duration systole.
Monitoring:
Monitor and evaluate the filling and emptying of the blood pump over sev-
eral cycles after each parameter change.
1. Adjust the
Rate
to generate sufficient blood flow.
2. Adjust
Filling
and
Emptying
. The goal is to completely fill and empty the blood
pump. A crease-free membrane in the end position indicates complete filling and
emptying. Complete filling and emptying ensures that the volume of the blood
pump is used and, therefore, that the required blood flow is achieved in the pa-
tient. The complete filling and emptying of the blood pump reduces the risk of de-
posits.
Ensure that the membrane moves continuously. At very high pressures, the
Movement of the membrane to the diastolic end position in order to fill the blood
pump.
1. Suction of blood
2. Diastolic end position
Emptying of the blood pump through ejection of the membrane.
3. Ejection of the membrane
4. Systolic end position
Содержание EXCOR Adult
Страница 2: ...2 1015002x00 en ...
Страница 12: ...12 1015002x00 Table of Contents en ...
Страница 14: ...14 1015002x00 Contact en ...
Страница 66: ...66 1015002x00 Driving Unit en ...
Страница 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Страница 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Страница 126: ...126 1015002x00 Replacing the Driving System en ...
Страница 128: ...128 1015002x00 Weaning and Explantation en ...
Страница 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Страница 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Страница 154: ...154 1015002x00 Manual Pump en ...
Страница 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Страница 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Страница 194: ...194 1015002x00 Technical Specifications en ...
Страница 198: ...198 1015002x00 Overview Parameters en ...
Страница 206: ...206 1015002x00 EMC en ...
Страница 208: ...208 1015002x00 Patient ID Labels en ...
Страница 210: ...Abbrevations 210 1015002x00 en ...
Страница 212: ...212 1015002x00 en ...