02/ 2016
BENNING MM 6-1/ 6-2
25
7.4.1 µA DC current range
(BENNING MM 6-1)
The input resistance is approx. 3 kΩ.
Overload protection: 1000 V
AC/DC
Measuring range OL indication
Resolution
Measurement accuracy
600.0 µA
660.0 µA
0.1 µA
± (1.0 % of r 5 digits)
7.5 AC current ranges
(BENNING MM 6-2)
Overload protection:
- 11 A (1000 V AC/ DC) fuse, 20 kA, rapid on 10 A input
Max. measuring time:
- 3 minutes at > 5 A (pause > 20 minutes)
- 30 secondes at > 10 A (pause > 10 minutes)
Measuring
range
OL indication
Resolution
Measurement accuracy
in frequency range 45 Hz - 500 Hz (sine)
6.000 A
6.600 A
0.001 A
± (1.5 % of r 5 digits)*
10.00 A
20.00 A
0.01 A
± (1.5 % of r 5 digits)*
* 6 A measuring range from ≥ 20 mA, 10 A measuring range from ≥ 100 mA
7.5.1 μA AC current range
(BENNING MM 6-1)
The input resistance is approx. 3 kΩ.
Overload protection: 1000 V
AC/DC
Measuring
range
OL indication
Resolution
Measurement accuracy
in frequency range 45 Hz - 500 Hz (sine)
600.0 µA
660.0 µA
0.1 µA
± (1.5 % of r 5 digits)*
* Measuring range from ≥ 1 µA
7.6 Resistance ranges
Overload protection: 1000
V
AC/ DC
Measuring range OL indication
Resolution
Measurement accuracy
600.0 Ω
660.0 Ω
0.1 Ω
± (0.9 % of r 8 digits)
6.000 kΩ
6.600 kΩ
0.001 kΩ
± (0.9 % of r 5 digits)
60.00 kΩ
66.00 kΩ
0.01 kΩ
± (0.9 % of r 5 digits)
600.0 kΩ
660.0 kΩ
0.1 kΩ
± (0.9 % of r 5 digits)
6.000 MΩ
6.000 MΩ
0.001 MΩ
± (0.9 % of r 5 digits)
40.00 MΩ
40.00 MΩ
0.01 MΩ
± (1.5 % of r 8 digits)*
* Measured values > 10 MΩ might cause changing values on the display
(max. ± 50 digits).
7.7 Diode testing
Overload protection: 1000 V
AC/ DC
Max. no-load voltage: 1.8 V
Measuring range OL indication
Resolution
Measurement accuracy
1.500 V
1.550 V
0.001 V
± (0.9 % of r 5 digits)
7.8 Continuity testing
Overload protection: 1000
V
AC/ DC
The built-in buzzer sounds in the case of a resistance R less than 20 Ω up to 200
Ω. For a resistance R higher than 200 Ω, the buzzer does not emit an acoustic
signal. Additionally, in case of continuity, the red LED
N
on the top side of the
device lights.
Measuring range OL indication
Resolution
Measurement accuracy
600.0 Ω
660.0 Ω
0.1 Ω
± (0.9 % of r 8 digits)
7.9 Capacity ranges
Conditions: capacitors discharged and connected in accordance with the polar-
ity stated.
Overload protection: 1000
V
AC/ DC
Measuring range OL indication
Resolution
Measurement accuracy
1.000 µF
1.100 µF
0.001 µF
± (1.9 % of r 8 digits)
10.00 µF
11.00 µF
0.01 µF
± (1.9 % of r 5 digits)