BENNING IT 130 

short instructions 

 

- 4 - 

 

Start-up guide

 

1.1  Safety and operational considerations 

The following symbols are used in the user manual and on the test equipment: 

 

  Important, danger, must comply with documentation! 

  Warning of electrical danger! 

  Protection class II 

   

  Ground (voltage to earth) 

 

 Warnings related to safety – general information 

 

 

This document is not a supplement to the Instruction manual! Please find the operating 

manual as PDF file on the enclosed CD-ROM. 

 

If the test equipment is used in a manner not specified in this user manual, the protection 

provided by the equipment could be impaired! 

 

Read this user manual carefully, otherwise the use of the instrument may be dangerous 

for the operator, the instrument or for the equipment under test! 

 

Do not use the instrument or any of the accessories if any damage is noticed! 

 

Consider all generally known precautions in order to avoid risk of electric shock while 

dealing with hazardous voltages! 

 

If  the  fuse  F1  has  blown,  please  follow  the  instructions  of  this  operating  manual  to 

replace  it!  Only  use  a  fuse  complying  with  the  specification  (see  chapter  7.1)  as 

replacement.  

If one of the fuses F2 or F3 has blown, the device must not be used anymore. In this case, 

the device must be sent to BENNING for inspection and repair. 

 

Do not use the instrument in AC supply systems with voltages higher than 550 V AC. 

 

Service,  repairs  or  adjustment  of  instruments  and  accessories  is  only  allowed  to  be 

carried out by a competent authorized personnel! 

 

Please  use  standard  or  optional  BENNING  accessories  only  which  are  available  from 

your authorized specialty retailer! 

 

Consider that protection category of some accessories is lower than of the instrument. 

Test tips and Tip "Commander" have removable caps. If they are removed the protection 

falls to CAT II. Check markings on accessories! 

cap off, 18 mm tip:  

CAT II up to 1000 V to earth 

cap on, 4 mm tip: 

CAT II 1000 V / CAT III 600 V / CAT IV 300 V to earth 

 

The instrument come  supplied with rechargeable NiMh battery  cells. The cells should 

only be replaced with the same type as defined on the battery compartment label or as 

described  in  this  manual.  Do  not  use  standard  alkaline  battery  cells  while  the  power 

supply adapter is connected, otherwise they may explode! 

 

Hazardous voltages  exist inside the instrument. Disconnect all test leads,  remove the 

power supply cable and switch off the instrument before removing battery compartment 

cover. 

 

Do not connect any voltage source on C1 input. It must be used only for connecting the 

current clamp adapters recommended by BENNING. Maximal input voltage is 3 V! 

 

All normal safety precautions must be taken in order to avoid risk of electric shock while 

working on electrical installations! 

Содержание 044103

Страница 1: ...Installation Tester BENNING IT 130 Short instructions GB Detailed instruction manual on enclosed CD Rom...

Страница 2: ...o KG M nsterstra e 135 137 D 46397 Bocholt Phone 49 0 2871 93 0 Fax 49 0 2871 93 429 www benning de duspol benning de 05 2020 BENNING This document must not be reproduced or used in any other form wit...

Страница 3: ...and fuses 11 1 10 Calibration and Service 12 1 11 Optional accessories 13 2 Measurements 14 2 1 Null balance compensation of the test cables 14 2 2 TRMS voltage V AC DC frequency and phase sequence r...

Страница 4: ...inspection and repair Do not use the instrument in AC supply systems with voltages higher than 550 V AC Service repairs or adjustment of instruments and accessories is only allowed to be carried out b...

Страница 5: ...isconnect all test cables accessories from the tester and from the installation and switch the tester off before opening the cover of the battery fuse compartment Dangerous voltages may be applied to...

Страница 6: ...ntrast 8 ON OFF Switches the instrument power on or off The instrument automatically turns off 15 minutes after the last key was pressed 9 HELP CAL Help function with connection diagrams press for app...

Страница 7: ...rying case item no 10008291 1 x Commander test probe switchable by means of TEST key item no 044155 1 x test cable with shock proof plug item no 10008295 1 x universal three wire test cable black blue...

Страница 8: ...N Battery indication Battery capacity indication Low battery The storage battery charge condition is too low to ensure correct measuring results Recharge the storage batteries or replace the batteries...

Страница 9: ...urrent clamp can be increased Measured signal is out of range clipped Results are impaired Simple error in the IT network Fuse F1 F2 or F3 is defective If one of the fuses F2 or F3 has blown the devic...

Страница 10: ...T network RCD TESTING according to EN 61008 EN 61009 IEC 60364 4 41 BS 7671 AS NZS 3017 SET ISC FACTOR setting 0 20 3 00 SELECT COMMANDER ON OFF INITIAL SETTINGS CLAMP SETTINGS BENNING CC 1 044037 BEN...

Страница 11: ...e batteries will be recharged automatically as soon as the charger is connected to the charging socket An integrated protective circuit controls the charging process Power supply socket polarity Note...

Страница 12: ...rantees compliance with the technical and accuracy specifications stated in the operating manual for the first 12 months after the delivery date To maintain accuracy of the measuring results the devic...

Страница 13: ...terior lighting Commander test plug For shock proof socket switchable with TEST and MEM keys with PASS FAIL indication by means of green red LED PE contact electrode for detecting a phase voltage at t...

Страница 14: ...t the test cables Shorted test leads 3 Press the key 4 Press the key CAL HELP After performing test leads compensation first measured value and then 0 00 is displayed Successful compensation is displa...

Страница 15: ...se adapter optional in a three phase system 3 The measuring result can be stored by means of the MEM key Uln Voltage between phase L and neutral conductors N Ulpe Voltage between phase L and protectiv...

Страница 16: ...500 V 1000 V Minimum limiting value without 0 01 M 200 M 3 Connection diagrams Connection of 3 wire test lead and Commander probe tip 4 Press and hold the key until result is stabilized Double click...

Страница 17: ...NUITY function 3 Connection diagrams Connection of the three wire test cable and of the optional measuring line BENNING TA 5 044039 Connection of 3 wire test lead and Commander probe tip 4 a R LOW low...

Страница 18: ...actor for test current x I N 1 2 5 x I N Limiting value for contact voltage 25 V 50 V 3 Connection diagrams Connecting the Commander test plug optional TN 044149 and the 3 wire test lead 4 Press the k...

Страница 19: ...CD Fuse type gL gG gG B C K D Nominal current of the fuse Tripping time of the fuse 3 Connection diagrams Connecting the Commander test plug 044149 and the 3 wire test lead 4 Press the key 5 The measu...

Страница 20: ...tion of the optional Commander test plug 044149 and of the three wire test cable Voltage drop Connection of the optional Commander test plug 044149 and of the three wire test cable 4 Press the key 5 T...

Страница 21: ...nce Maximum limiting value without 1 5 k 3 Connection diagrams Connection of the optional earthing set 044113 4 Press the key 5 The measuring result can be stored by means of the MEM key R Earth resis...

Страница 22: ...unction I Note Please set the current clamp adapter used under SETTINGS 2 Connection diagrams Connection of the optional current clamp adapter BENNING CC 1 044037 BENNING CC 3 044038 4 Press the key 5...

Страница 23: ...Set function 2 Set limit LUX Minimum illumination OFF 0 1 lux 20 klux 3 Connection diagrams Positioning of the luminous intensity sensor BENNING luxmeter type B 044111 4 Press the key 5 The measuring...

Страница 24: ...t RISO ISFL Single fault leakage current only in IT networks Maximum limiting value without 3 0 mA 20 0 mA 3 Connection diagrams Connection of the three wire test cable 4 Press the key 5 The measuring...

Отзывы: