Operation
49
MK2430_D00129_02_M_XXEN/09.2019
7.2 Quick-start guide
The illustrations below serve as examples.
7.2.1 Display in fault-free operating condition
There are no warnings or alarms pending.
The green "Normal" LED is lit.
The LC display shows the programmed standard display.
A maximum of 3 measured values can be displayed in lines 1…3.
– Line 1…3:
user-definable standard display text.
– Line 4
status bar, indicates the time of day (can be switched off).
7.2.2 Display in fault condition
A warning or an alarm message is pending.
Depending on the type of fault, either the yellow "Warning" LED or the red
"Alarm" LED will light up. The green "Normal" LED goes out.
At the same time the buzzer sounds. If the cause of fault cannot be remedied
immediately, the buzzer can be muted by pressing the "
" (5) button.
The LC display shows details about the message.
– Line 1:
Standard display: "System:"
User-defined text here: "Intensive care unit 03"
– Line 2…3:
Message text, with measured value (if applicable)
Power supply: OK
Med. gases: OK
09:50
Intensive care unit 03
Insulation fault
Measured value 43 k
Ω
xx/yy zzz 09:50