38
www.bempu.com
on power
(60% dip
(60% dip
during power
supply
in UT)
in UT)
mains interruptions, it is
input lines
for 5
for 5
recommended that the
IEC
70% UT
70% UT
device be powered from
61000-4-11
(30% dip
(30% dip
an uninterruptible
in UT)
in UT)
power source.
for 25
for 25
cycles
cycles
<5% UT
<5% UT
(>95% dip
(>95% dip
in UT)
in UT)
for 5 sec
for 5 sec
Power
3 A/m
3 A/m
Power frequency mag-
frequency
netic fields should be
(50/60Hz)
at levels characteristic
magnetic
of a typical location in a
field
typical commercial or
IEC
hospital environment.
61000-4-8
Conducted
3 Vrms
3 Vrms
Portable and mobile RF
RF
150 kHz to
150 kHz to
communications equip-
IEC
80 MHz
80 MHz
ment should be used no
61000-4-6
closer to any part of the
device, including ca-
bles, than the recom-
mended separation
distance calculated
from the equation ap-
plicable to the frequency
of the transmitter.
Radiated
3 V/m
3 V/m
Recommended separa-
RF
80 MHz to
80 MHz to
tion distance
IEC
2.5 GHz
2.5 GHz
61000-4-3
d = 0.35 √P
d = 0.35 √P 80 MHz to
800 MHz
d = 0.70 √P 800 MHz to
2.5 GHz
Where (P) is the maxi-
mum output power rating
Содержание ApneBoot