a) Probe Not Connected
It indicates that the pulse oximeter probe cable is not connected to the main
unit.
b) Stimulator Not found
It indicates that the stimulator boot cable is not connected to the main unit.
Step 3:
Connect the device to the infant.
(i) Placement of Pulse Oximeter Probe: Place the Pulse Oximeter Probe on
the infant’s foot or other peripheral body site, suitable for measuring pulse
oximeter based oxygen saturation and pulse rate.
(ii) Placement of the Stimulator Boot: The stimulator boot must be wrapped
against the foot sole of the infant snuggly, neither too tight nor too loose.
Ensure that the stimulator wire passes through the toe side and not through
the heel side.
WARNING:
The stimulator boot should ONLY be placed on the foot
sole and not on any other body site.
WARNING:
Do not put the stimulator boot if the infant’s foot has any
skin allergies and is chapped/itchy/red. This may result in aggravation of the
allergy.
WARNING:
Clean the probes using surgical spirit and cotton swab
before placing it against the infant’s foot sole.
Refer
to cleaning instructions
in Section 5.1.
The pulse oximeter probe will start measuring the infant’s Pulse (PR) and
Oxygen Saturation (SPO
2
). Make sure the device continuously displays both
the values on the screen of the unit. If not, check the placement of the Pulse
Oximeter Probe.
2.4 Method of Probe Placement
(i) Stimulator Boot
Place the stimulator boot on the infant’s foot sole with toes pointing in the
direction of wire.
Wrap the strap (a) over the foot in the direction marked as 1 in the below im
-
age. Lock the strap (a), so that there is a firm fitting of the foot.
!
!
!