20
9.Troubleshooting
9.Troubleshooting
Problems
Causes
Solutions
Skin feels a painful
burning sensation,
or the stimulation
becomes weak
-
ened.
Adhesive gel pads
are not adhering
firmly to the skin.
The gel pads are
too dry.
Apply a few drops of wa-
ter to the adhesive surface
of each pad and make
sure the pads are pressed
firmly to the skin during
application.
Unit is on, but no
sensation is felt in
the pads.
Both pads are not
firmly on the skin.
There is a loose
connection.
The intensity
level needs to be
increased.
Make sure all pads are
firmly pressed to the skin.
Make sure all connections
are well from the unit to
the wires and the wires to
the pads.
Increase the intensity level.
Adhesive gel pads
do not stick to skin
even after cleaning
and moistening
the gel pad.
Adhesive gel
pads need to be
replaced.
Replace the pads. Pads
can be purchased from
Belifu official store or
contact support@belifu.
com.
Unit does not turn
on or the screen is
dim.
Battery is low.
Recharge the battery.
Does not provide
adequate pain
relief.
Not using it long
enough.
Use your massager 20-
30 minutes at a time, 3-6
times per day.
Unit turns on, all
of the modes and
channels flash. Can
not select a mode
and adjust the
intensity.
Does not work.
Always use two
pads (one chan-
nel) or four pads
(two channels) at
the same time as
it will not work
with just one pad
or three pads.
When you use two pads
(one channel) or four
pads (two channels) at
the same time, make sure
you have placed all pads
on your body, otherwise it
won't work.
The pluses strong
on one side and
weak on the other
side.
Different nerve
distribution in
different parts of
the body, different
skin sensitivity, or
insufficient skin
contact with the
electrodes.
Please wipe the skin with
a wet towel before use to
ensure adequate contact
with the electrode. It
suggested that the two
electrodes should be sym-
metrical when attached to
the body.