RF emissions
CISPR 11
Class B
The FDES115 suitable for use
in all establishments, including
domestic establishments and
those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emis-
sions
IEC 61000-3-2
Class A
Voltage fluctua-
tions
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic
immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic
immunity
The FDES115 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the FDES115
should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compli
-
ance level
Electromagnetic envi-
ronment -guidance
Electrostat-
ic
discharge
(ESD)
IEC 61000-
4-2
± 8 kV
contact
± 2 kV, ± 4
kV,
± 8 kV,
± 15 kV air
± 8 kV
contact
± 2 kV, ± 4
kV, ± 8 kV,
± 15 kV air
Floors should be
wood, concrete or ce
-
ramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrostat-
ic
transient /
burst
IEC 61000-
4-4
± 2 kV for
power
supply lines
± 1 kV for
input/output
lines
± 2 kV for
power
supply
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
surge
IEC 61000-
4-5
±
1 kV differ-
ential mode
±
2 kV com-
mon mode
± 1 kV
differential
mode
Mains power quality
should be that of
atypical commercial
or hospital environ
-
ment.