background image

32

8.    OPERATION AS A CLOCK

When configured as a clock the BA578E can display
local time in a variety of twelve or twenty four hour
formats.

The displayed time can be adjusted via

the front panel push buttons which may be protected
by a user definable four digit security code to
prevent unauthorised or accidental adjustment.

The clock may be synchronised to an external time
standard via the instrument's reset terminals. When
these two terminals are connected together by an
external switch contact, the clock display will be
reset to a preconfigured time and will resume
running from this time when the contacts are
opened.

The optically isolated status open collector output
can be used to monitor the clock or to perform
simple control functions. 

When fitted with the optional galvanically isolated
control outputs the clock can be configured to turn
each solid state output

on

and

off

twice during each

twenty four hour period.

Fig 13 shows a simplified block diagram of the
BA578E configured as a clock.

The instrument can be supplied with the following
factory fitted accessories:

Backlight Internally powered

Dual isolated control outputs

Fig 13

BA578E block diagram with clock 
configuration.

8.1

Initialisation and loss of power

Each time power is applied to a BA578E
initialisation is performed. After a short delay the
following display sequence occurs:

All segments of the display are activated

Instrument starts functioning, using the
configuration information stored in the
instrument's permanent memory.

Following initialisation, the instrument will have a
flashing display which is a request for the local time
to be entered.

The clock will not start to function

until a display time has been entered, or the remote
reset

contacts

are

closed

and

opened

to

synchronise the clock to a preconfigured time.

If during normal operation the power supply is
interrupted for more than 30ms, the display will
return to the flashing condition and the display time
will have to be re-entered.

This time may be

increased by powering the Clock from a lightly
loaded

instrument

supply

with

large

output

capacitance.

8.2

Controls when configured as a clock 

The BA578E clock is configured and adjusted via
four front panel push buttons. In the display mode
i.e. when the instrument is displaying time, the push
button functions are:

Push Button Functions

(

 + 

&

  

Shows firmware version

(

 + 

 

Access to configuration menu

Note:

When optional control outputs are fitted, the

BA578E clock may be configured to provide direct
access to the control outputs from the display mode
when the

(

and

*

push buttons are operated

simultaneously. - see section 8.4.16

8.3

Displays when configured as a clock

The BA578E clock has a single digital display plus
annunciators.
 

Time

Shows time in selected 12 or

display

24 hour format.

Reset

Activated while clock is being

annunciator

synchronised and external
reset contacts are closed.

Status output

RTx shown while status  

annunciator

output is activated. 

Control output

Show status of both optional

annunciators

control outputs.

Содержание BA578E

Страница 1: ...Issue 4 3rd July 2018 BA578E Two input General purpose Timer or Clock Issue 4...

Страница 2: ...14 Cycle count cycl cnt 5 5 15 Restart delay r5t dela 5 5 16 Adjusting the set time 5Et t from the display mode AC5Et t 5 5 17 Direction of count uP or dn 5 5 18 Power fail p fail 5 5 19 Local reset...

Страница 3: ...ff 1 5tat on 2 5tat off 2 8 4 10 Control output 1 oP1 8 4 11 Enable Control output 1 EnbL 8 4 12 Control output 1 on and off times oP 1 on 1 oP 1 off 1 oP 1 on 2 oP 1 off 2 8 4 13 Control output 2 oP2...

Страница 4: ...4...

Страница 5: ...nstrument 2 SYSTEM DESIGN Fig 1 illustrates the basic circuit that is used for all BA578E Timer installations BA578E Clock installations are the same except that the two input terminals are not used F...

Страница 6: ...input ranges are independently selectable for each input in the BA578E Timer configuration menu VoLt5 L and VoLt5 H allowing almost any sensor with a voltage output to be used The BA578E contains a se...

Страница 7: ...int between the instrument enclosure and the panel the rear of the instrument has IP20 ingress protection The BA578E may be installed in any panel providing that the operating temperature is between 4...

Страница 8: ...p body d Place a clamp in the recess on each side of the instrument pulling gently to slide it onto the dovetail as shown in Fig 6 Push the knurled screw slightly forward to engage the thread and tigh...

Страница 9: ...t at the rear of the instrument 4 ACCESSORIES 4 1 Display backlight The BA578E Timer or Clock can be supplied with a factory fitted backlight that produce green illumination enhancing display contrast...

Страница 10: ...ion New instruments can be supplied with a tag number or application information printed onto the rear panel adjacent to the terminals This information is not visible from the front of the instrument...

Страница 11: ...ntained by the instrument which can be viewed by operating the front panel push buttons and reset from the display mode or from within the configuration menu Fig 8 shows a simplified block diagram of...

Страница 12: ...d to the restart delay r5t dela See 6 5 16 Shows firmware version Accesses the configuration menu 5 3 Displays when configured as a Timer The BA578E has two digital displays and associated annunciator...

Страница 13: ...the Pause state by a stop event which could be an input on input b or operation of the button Reset In this state the Timer is readied for operation The Timer is stopped and loaded with zero for timi...

Страница 14: ...figuration functions Throughout this manual front panel push buttons are shown as and Legends displayed by the instrument are shown in a seven segment font just as they appear on the instrument e g in...

Страница 15: ...l of input debounce applied to the input A to prevent false counting dEFAuLt HEAVY LiGHt See section 5 5 6 Display Summary of function inPut b Configuration of input b As configuration of inPut A Ener...

Страница 16: ...protect access to 5Et t from the display mode See section 5 5 16 Display Summary of function uP or dn Direction of count Defines whether the Timer times up from zero to the set time 5Et t or times do...

Страница 17: ...ary of function Control output 1 on oP1 on Control output 1 turns on when the Timer enters the state selected in this function Output phase can be reversed See section 5 5 30 Control output 1 off oP1...

Страница 18: ...on menu One of following six types of input may be selected Display Input type Switching thresholds Low High oPCoL Open collector 2 10k VoLt5 L Voltage pulse low 1 3V VoLt5 H Voltage pulse high 3 10V...

Страница 19: ...disabled 5 5 9 Starting stopping the Timer 5tAr5toP The Timer may be started and stopped by signals on Input A and Input b or by operation of the front panel or push buttons To check or change the con...

Страница 20: ...20...

Страница 21: ...21...

Страница 22: ...eveal the number of times the timer cycle is to be repeated with the most significant flashing The flashing digit may be adjusted by pressing the or button followed by the button to transfer control t...

Страница 23: ...on the upper display Timing resumes when a start instruction is received If a start input exists when power is restored timing will start immediately Continue When power is restored the Timer will con...

Страница 24: ...tates such as n re5et If the status output is required to be on in only one Timer state this can be achieved with the following configuration 5tat on state required e g run 5tat off n state required e...

Страница 25: ...to the status output sub menu 5 5 28 Control output 1 optional op1 Control output 1 is an optional factory fitted galvanically isolated solid state switch contact which can be configured to turn on an...

Страница 26: ...ions Display Control output 1 turns off when Timer enters selected state re5et Reset state run Running state pau5ed Paused state delay Restart delay state done Complete state Control output 1 turns of...

Страница 27: ...ed to gain access New instruments are configured with the default security code 0000 which allows unrestricted access to all configuration functions To enter a new security code select CodE from the c...

Страница 28: ...option and should be specified when the instrument is ordered Control output 1 is used to switch the intrinsically safe solenoid valve and control output 2 is wired to the Timer s reset terminals to p...

Страница 29: ...uP or dn dn Power failure p fail idle Local total reset rE5Et EnbL on Local grand total reset CLr Gtot oFF External reset e r5et fa5t Enable control output 1 EnbL on Control output 1 on at oP1 on run...

Страница 30: ...s in the configuration menu Unable to enter configuration menu Incorrect security code That the correct security code is being used Contact BEKA if code is lost 7 2 Fault finding after commissioning E...

Страница 31: ...regularly checked Inspection frequency should be chosen to suit the environmental conditions 7 5 Guarantee Instruments which fail within the guarantee period should be returned to BEKA associates or o...

Страница 32: ...t starts functioning using the configuration information stored in the instrument s permanent memory Following initialisation the instrument will have a flashing display which is a request for the loc...

Страница 33: ...ion functions Throughout this manual push buttons are shown as and and legends displayed by the Clock are shown in a seven segment font just as they appear on the instrument e g di5PLAY and CodE Acces...

Страница 34: ...34...

Страница 35: ...35...

Страница 36: ...output can turn on open collector on and off open collector off once or twice in each twenty four hour period This function contains five sub function EnbL Enables Status output and defines if it turn...

Страница 37: ...to the display mode To configure the Clock enter the configuration menu by pressing the and buttons simultaneously until function is displayed 8 4 4 Display format di5PLAY The BA578E Clock may be con...

Страница 38: ...ed on 1 Turns on off once per 24 hours on 2 Turns on off twice per 24 hours When the required option is displayed operating will enter the selection and return to the EnbL prompt from which another su...

Страница 39: ...en entered return to the configuration menu by pressing the button twice 8 4 13 Control output 2 oP2 Control output 2 is an optional factory fitted output which can be configured to turn on output clo...

Страница 40: ...mes from the display mode may be protected by a four digit alphanumeric security code which must be entered to gain access Default security code 0000 allows unrestricted access With ACCd displayed pre...

Страница 41: ...n change the instrument to a Clock and return the display to FunCtion in the configuration menu See 8 4 3 Step 3 Select display format Using the or button select di5PLAY in the configuration menu and...

Страница 42: ...ess to the configuration menu Using the and buttons select CodE from the configuration menu and press which will reveal 0000 with the first digit flashing This example requires the security code to be...

Страница 43: ...If a BA578E fails after it has been functioning correctly as a Clock the following table may help to identify the cause of the failure Symptom Cause Check No display No power supply That there is bet...

Страница 44: ...4 10 6 Customer comments BEKA is always pleased to receive comments from customers about our products and services All communications are acknowledged and whenever possible suggestions are implemented...

Отзывы: