It is preferable to make this substitution as close to
the RTD as possible, but if necessary the
transmitter may be calibrated in a workshop.
Select '4rtd' from the 'Cond' sub-menu and press 

P.

The transmitter will display scrolling decimal points
while the reference is being updated followed by
'Ent' prior to returning to the '4rtd' prompt.  This
completes adjustment of this reference.

5.8.5  Output current reference   'OP'

Warning: ensure plant safety

before performing this adjustment

This adjustment is only necessary if the transmitter
4/20mA output current is used.   A method of
accurately measuring the transmitter output current,
such as DVM, is required to calibrate this
reference.   

Select 'OP' from the 'Cond' sub-menu and press 

P

,

the transmitter will display 'SurE'. If it is safe to
change the transmitter output current between 4
and 20mA continue by pressing

 P

.  The transmitter

will display 'ZEro', pressing

 P 

again will change the

display to '.004A' and the transmitter output current
 to approximately 4mA. The transmitter output
current shown on the external current measuring
instrument should be adjusted to 4.000 ±0.001mA
using the transmitter 

Up

 and 

Down

 buttons.  When

correct press the

  E

 push-button to return to the

'ZEro' prompt.   

To adjust the 20mA output current reference select
the 'SPAn' prompt by operating the 

Up

 or 

Down

button.  Pressing 

P

 will change the display to

'.020A' and the transmitter output current to
approximately 20mA.  The transmitter output
current shown on the external current measuring
instrument should be adjusted to 20.000± 0.001mA
using the transmitter  

Up

 and 

Down

 buttons.  When

correct press the 

E

 push-button twice to return to

the 'OP' prompt.  This completes adjustment of this
reference.

6.  TRIM MENU

The Trim Menu enables the BA574C transmitter to
be adjusted using an external voltage, resistance or
temperature calibrator.  For maximum accuracy the
complete loop including the thermocouple or
resistance thermometer may be calibrated by
heating or cooling the primary element. 

Before using the Trim Menu all the functions in the
'InPut' and 'CAL' sections of the programme menu
must be defined, including the 4/20mA output
range, and for voltage inputs the input and display
ranges.

If a voltage calibrator is used to simulate a
thermocouple the transmitter cold junction
compensation must be turned off.  Alternatively, the
temperature at the BA574C input terminals may be
measured and the  voltage calibrator corrected for
ambient temperature, although this is likely to
produce less accurate results.

6.1  Calibration  procedure using Trim Menu

The structure of the Trim Menu is shown in Fig 7.
Access to the menu is obtained by operating the 

P

and 

Down

 push-buttons simultaneously.  If the

transmitter is not protected by a security code the
first parameter 'ZEro' will be displayed.  If the
transmitter is protected by a security code, 'COdE'
will be displayed first.  Pressing 

P

 again will enable

the security code to be entered digit by digit, using
the 

Up

 and 

Down

 buttons to change the flashing

digit, and the 

P

  push-button to move to the next

digit.   If the correct code is entered pressing 

E

 will

cause the first parameter 'ZEro' to be displayed.   
If an incorrect code is entered, or if no button is
pressed for ten seconds, the transmitter will
automatically return to the display mode.

Fig 7 Trim Menu

The 'ZEro' sub-menu allows the transmitter display
and output current to be independently trimmed at
the bottom of the input range.  To make the adjust-
ment connect a calibrator to the transmitter input
terminals to simulate minimum input, or place the
primary element in a vessel at the minimum
temperature. With 'ZEro' displayed, pressing the 

P

push-button will cause the temperature display to
return. 

      16

Содержание BA574C

Страница 1: ...BA574C field mounting indicating temperature transmitter issue 4 Issue 4 26th September 2008...

Страница 2: ...Voltage input reference 5 8 3 3 wire RTD reference 5 8 4 4 wire RTD reference 5 8 5 Output current reference 6 Trim menu 6 1 Calibration procedure using Trim menu 6 2 Periodic re calibration using Tr...

Страница 3: ...n 5 you will find information on setting up for Section 5 3 Themocouple input Section 5 4 Resistance thermometer input Section 5 5 Voltage inputs A word of warning Section 5 8 explains how you can cal...

Страница 4: ...4...

Страница 5: ...and controls the 4 20mA output current The transmitter and the optional alarms are powered from the 4 20mA current loop but the optional display backlight requires a separate supply Each time power is...

Страница 6: ...must have an output greater than 15 2V 4 INSTALLATION 4 1 Location The BA574C indicating transmitter can be supplied in either a glass reinforced polyester GRP or an epoxy painted aluminium enclosure...

Страница 7: ...ation to be performed against the internal transmitter refer ences The Trim Menu enables the transmitter display and the 4 20mA output current to be calibrated against an external standard such as a v...

Страница 8: ...ermocouple and resistance thermometer inputs The BA574C transmitter will always display sensor temperature Programming allows the display to be in degrees Centigrade or Fahren heit and for the display...

Страница 9: ...the output is 4mA and SPAn the display at which the output is 20mA See section 5 6 COdE Security code Defines a four digit numeric code which must be entered to gain access to the programme and the t...

Страница 10: ...10...

Страница 11: ...11...

Страница 12: ...is only available when the transmitter is displaying temperature in o C When o F is selected the resolution is always one degree To check or change the display resolution select rESn from the sub menu...

Страница 13: ...l point can be positioned between any of the display digits or it may be absent After selecting voltage input from the InPut menu press P to reveal the d P sub menu Press P again to display the curren...

Страница 14: ...ay can be changed by pressing the Up or Down buttons When the first digit is correct pressing P will transfer control to the next digit When the least significant digit has been adjusted press E to re...

Страница 15: ...is a ten second time out on password entry Pressing E will then give access to the sub menu containing the four refences 5 8 2 Voltage input reference U IP With the BA574C input terminals connected t...

Страница 16: ...nsmitter to be adjusted using an external voltage resistance or temperature calibrator For maximum accuracy the complete loop including the thermocouple or resistance thermometer may be calibrated by...

Страница 17: ...shown on the DVM or other instrument When correct press the E button to store the trim correction in perma nent memory and return the BA574C transmitter to the SPAn prompt This completes re calibratio...

Страница 18: ...u Step 8 Input open circuit drive Using the Up or Down button select burn from the sub menu and press P to reveal the current status Select uP using the Up or Down button and press E three times to re...

Страница 19: ...sh P which will cause the transmitter to display the thermocouple temperature Adjust the display with the Up and Down buttons until it is 240 0 Note Immediately the Up or Down button is operated the a...

Страница 20: ...ammable on site a single spare transmitter or electronic assembly may therefore be used to replace any instrument which fails We recommend that suspect instruments are returned to BEKA associates or o...

Страница 21: ...nt panel 10 0 ALARMS The BA574C can be supplied with two solid state single pole alarm outputs Each output may be programmed as a high or low alarm with a normally open or a normally closed output Fig...

Страница 22: ...he operating mode See section 10 3 2 HI LO Alarm function Defines whether the alarm has a high or low function See section 10 3 3 no nc Normally open or normally closed output Determines whether the s...

Страница 23: ...23 Fig 11 Programme structure for optional alarms...

Страница 24: ...m output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc p...

Страница 25: ...e has been entered press E twice to return to the AcSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons si...

Страница 26: ...tter is installed in a poorly illuminated area The backlight is segregated from the measuring circuit and may be powered from any supply with an output between 18 and 30V dc as shown in Fig 13 If the...

Отзывы: