background image

5.8  Calibrating the internal references

WARNING!

Changing the internal references will
alter the instrument performance. Do
not not attempt to recalibrate these
references unless you have adequate
test  equipment and have read the
following instructions. For routine
calibration of the the instrument use the
Trim Menu described in section 6.

If the internal references are used for calibrating
the BA574C transmitter they may require periodic
checking against external standards. Annual checks
will produce acceptable performance, although less
frequent checks are usually adequate.  

If the BA574C transmitter is calibrated against an
external voltage, resistance or temperature
calibrator using the Trim Menu, the internal
references should never need re-calibrating.
Details  of  the  Trim  Menu  are  contained  in
section 6.
 
There are four internal references:

Voltage input
3-wire RTD
4-wire RTD
Current output  

The input voltage reference must be adjusted first
irrespective of the transmitter application, but  it is
only necessary to adjust the other three references
if the function to which they relate is being used.
For example, a BA574C with a thermocouple input
from which the 4/20mA output current is being used
would require the voltage input and the current
output reference adjusting.

5.8.1 Conditioning internal references:  'Cond'

WARNING!

Changing the internal references will
alter the instrument performance. Do
not not attempt to recalibrate these
references unless you have adequate
test  equipment and have read the
following instructions. For routine
calibration of the the instrument use the
Trim Menu described in section 6.

These functions are contained in a protected
sub-menu to minimise the possibility of accidental
adjustment.  Select 'Cond' from the menu and press

P

.  The transmitter will display '0000'. 

Using the 

Up

 and 

Down

 buttons and the 

P

 button to

move to the next digit enter the password 'SurE.
(Note that 'S' is entered as '5')  There is a ten
second time-out on password entry. Pressing 

E

 will

then give access to the sub-menu containing the
four refences.

5.8.2  Voltage input reference  'U.IP'

With the BA574C input terminals connected to an
accurate voltage calibrator, select 'U.IP' from the
'Cond' sub-menu and press 

P.

  The transmitter will

respond by displaying 'Lo' which is a request for an
accurate -75.000mV input.  When the input is
correct press 

P

  to store this figure in permanent

memory.  The transmitter will display scrolling
decimal points while the reference is being updated
and will then return to the 'Lo' prompt.

Select 'Hi' from the sub-menu which is a request for
an ac75.000mV input.  When the input is
correct press 

P

  to store this figure in permanent

memory.  The transmitter will display scrolling
decimal points while the reference is being updated
and will then return to the 'Hi' prompt.   This
completes adjustment of the voltage input
reference.

5.8.3  3-wire RTD reference  '3rtd'

This adjustment, which is only necessary if the
transmitter is used with a three wire resistance
thermometer, should be performed after the
voltage reference has been calibrated. To adjust
the 3-wire RTD reference the resistance
thermometer should be replaced by a 100 ohm
standard resistor or calibrator, with the
compensation loop  remaining connected between
terminals 3 and 4.   Although it is preferable to
make this substitution as close to the RTD as
possible, transmitters may be calibrated in a
workshop.

Select '3rtd' from the 'Cond' sub-menu and press 

P.

The transmitter will display scrolling decimal points
while the reference is being updated followed by
'Ent' prior to returning to the '3rtd' prompt.  This
completes adjustment of this reference.

5.8.4  4-wire RTD reference  '4rtd'

This adjustment, which is only necessary if the
transmitter is used with a four wire resistance
thermometer, should be performed after the
voltage reference has been calibrated.  To adjust
the 4-wire RTD reference the resistance
thermometer should be replaced by a 100 ohm
standard resistor or calibrator, with both
compensation loops from terminals 1 and 4
connected at the standard resistor or calibrator.

      15

Содержание BA574C

Страница 1: ...BA574C field mounting indicating temperature transmitter issue 4 Issue 4 26th September 2008...

Страница 2: ...Voltage input reference 5 8 3 3 wire RTD reference 5 8 4 4 wire RTD reference 5 8 5 Output current reference 6 Trim menu 6 1 Calibration procedure using Trim menu 6 2 Periodic re calibration using Tr...

Страница 3: ...n 5 you will find information on setting up for Section 5 3 Themocouple input Section 5 4 Resistance thermometer input Section 5 5 Voltage inputs A word of warning Section 5 8 explains how you can cal...

Страница 4: ...4...

Страница 5: ...and controls the 4 20mA output current The transmitter and the optional alarms are powered from the 4 20mA current loop but the optional display backlight requires a separate supply Each time power is...

Страница 6: ...must have an output greater than 15 2V 4 INSTALLATION 4 1 Location The BA574C indicating transmitter can be supplied in either a glass reinforced polyester GRP or an epoxy painted aluminium enclosure...

Страница 7: ...ation to be performed against the internal transmitter refer ences The Trim Menu enables the transmitter display and the 4 20mA output current to be calibrated against an external standard such as a v...

Страница 8: ...ermocouple and resistance thermometer inputs The BA574C transmitter will always display sensor temperature Programming allows the display to be in degrees Centigrade or Fahren heit and for the display...

Страница 9: ...the output is 4mA and SPAn the display at which the output is 20mA See section 5 6 COdE Security code Defines a four digit numeric code which must be entered to gain access to the programme and the t...

Страница 10: ...10...

Страница 11: ...11...

Страница 12: ...is only available when the transmitter is displaying temperature in o C When o F is selected the resolution is always one degree To check or change the display resolution select rESn from the sub menu...

Страница 13: ...l point can be positioned between any of the display digits or it may be absent After selecting voltage input from the InPut menu press P to reveal the d P sub menu Press P again to display the curren...

Страница 14: ...ay can be changed by pressing the Up or Down buttons When the first digit is correct pressing P will transfer control to the next digit When the least significant digit has been adjusted press E to re...

Страница 15: ...is a ten second time out on password entry Pressing E will then give access to the sub menu containing the four refences 5 8 2 Voltage input reference U IP With the BA574C input terminals connected t...

Страница 16: ...nsmitter to be adjusted using an external voltage resistance or temperature calibrator For maximum accuracy the complete loop including the thermocouple or resistance thermometer may be calibrated by...

Страница 17: ...shown on the DVM or other instrument When correct press the E button to store the trim correction in perma nent memory and return the BA574C transmitter to the SPAn prompt This completes re calibratio...

Страница 18: ...u Step 8 Input open circuit drive Using the Up or Down button select burn from the sub menu and press P to reveal the current status Select uP using the Up or Down button and press E three times to re...

Страница 19: ...sh P which will cause the transmitter to display the thermocouple temperature Adjust the display with the Up and Down buttons until it is 240 0 Note Immediately the Up or Down button is operated the a...

Страница 20: ...ammable on site a single spare transmitter or electronic assembly may therefore be used to replace any instrument which fails We recommend that suspect instruments are returned to BEKA associates or o...

Страница 21: ...nt panel 10 0 ALARMS The BA574C can be supplied with two solid state single pole alarm outputs Each output may be programmed as a high or low alarm with a normally open or a normally closed output Fig...

Страница 22: ...he operating mode See section 10 3 2 HI LO Alarm function Defines whether the alarm has a high or low function See section 10 3 3 no nc Normally open or normally closed output Determines whether the s...

Страница 23: ...23 Fig 11 Programme structure for optional alarms...

Страница 24: ...m output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc p...

Страница 25: ...e has been entered press E twice to return to the AcSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons si...

Страница 26: ...tter is installed in a poorly illuminated area The backlight is segregated from the measuring circuit and may be powered from any supply with an output between 18 and 30V dc as shown in Fig 13 If the...

Отзывы: