Step 5

Position dummy decimal point

Select 'd.P' from the main menu and
then press 

P

. Using the 

Up

 and 

Down

push-buttons position the dummy
decimal point in front of the least
significant digit.
Press 

E

 to return to the main menu.

Step 6

Calibrate display

With any input current between 4 and
20mA select 'SEt' from the main menu
and press 

P

. The indicator will display

'ZErO' in the sub-menu, press 

P

 to

reveal the existing zero display. Using
the 

Up

Down

 and 

P

 push-buttons

enter  the  required  zero  display  of
-50.0
Press 

E

 to return to the 'ZErO' prompt.

With any input current between 4 and
20mA press the Up push-button and
the indicator will display 'SPAn'. Press

P

 to reveal the existing span display.

Using the 

Up

Down

 and 

P

 push-

buttons enter the required span
display of 1050.0 Press 

E

 to return to

the 'SPAn' prompt.
Press 

E

 again to return to the main

menu.

Step 7

Return to display mode

Return to the display mode by
pressing 

'E'

.

7.  Maintenance

7.1 Fault finding during commissioning

If a BA527C or BA528C fails to function during
commissioning the following procedure should be
followed:

Symptom

  Cause

        Solution

No display

Incorrect

There should be 1V

wiring

between terminals 1 &
3 with terminal 1
positive.

No display,

Incorrect

Check that a current

and 0V

wiring or

is flowing in the loop.

between

no power

terminals

supply.

1 and 3.

Insufficient

Check supply voltage

loop voltage

and voltage drops

to operate

caused by all

indicator

components in the loop.

Indicator

Positive

The indicator has been

displays 1

over-range

incorrectly calibrated &
is trying to display a
number greater than
19999.

Indicator

Negative

The indicator has been

displays –1

0ver-range

incorrectly calibrated &
is trying to display a
number less than
-19999.

Unstable

4/20mA input

Check loop supply

display

has a large

voltage.

ripple content.

Unable to

Incorrect

Enter correct security

enter the

security

code or fit security

programme

code

link in override position.

mode

entered.

See Fig 7.

7.2 Fault finding after commissioning

ENSURE PLANT SAFETY BEFORE

STARTING MAINTENANCE

If a BA527C or BA528C fails after it has been
functioning correctly, the following procedure
should be followed:

Содержание BA527C

Страница 1: ...Issue 3 5th March 2009 BA527C BA528C loop powered 4 digit panel mounting indicators Issue 3...

Страница 2: ...urce 6 2 Using the internal calibrator 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comme...

Страница 3: ...o the 4 20mA input current provides the input signal for the analogue to digital converter Fig 1 Simplified block diagram Each time a 4 20mA current is applied to the instrument initialisation is perf...

Страница 4: ...any 4 20mA signal to provide a remote indication Fig 3 shows a typical application in which the output from a gas analyser drives a BA527C or BA528C indicator Again it is necessary to ensure that the...

Страница 5: ...trument and tighten until the indicator is secure as shown in Fig 5 Four clips are required to achieve an IP65 between a BA528C and the instrument panel Fig 4B BA528C dimensions c Connect the panel wi...

Страница 6: ...this manual push buttons are shown in italics e g P or Up push button and legends displayed by the indicator are shown within inverted commas e g CAL and ALr2 Access to the programme menu is obtained...

Страница 7: ...0mA loop See section 5 6 Display Description of function Cond Indicator conditioning This function provides access to a sub menu enabling the internal references to be calibrated and the mains line fr...

Страница 8: ...8...

Страница 9: ...adjusted press E to enter the new zero The indicator will display Ent for a few seconds while the information is being stored in memory and will then return to the ZErO prompt To adjust the display at...

Страница 10: ...alibrator to 20 000mA and again press P The indicator will display Ent when the 20mA reference has been updated and will then return to the 020A prompt Two operations of the E button will return the i...

Страница 11: ...requency of the mains supply is 50Hz and the existing security code is 1209 In the operating mode the indicator is required to display the input current as a percentage of span when the P push button...

Страница 12: ...ain menu Step 8 Return to the display mode Return to the display mode by pressing E 6 2 Using the internal calibrator As in 6 1 the BA527C is required to display 50 0 with a 4mA input 1050 0 with a 20...

Страница 13: ...ing If a BA527C or BA528C fails to function during commissioning the following procedure should be followed Symptom Cause Solution No display Incorrect There should be 1V wiring between terminals 1 3...

Страница 14: ...d 7 6 Customer comments BEKA associates is always pleased to receive comments from customers about our products and services All communications are acknowledged and whenever possible suggestions are i...

Страница 15: ...application of alarms Summary of programmable alarm functions Display Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section...

Страница 16: ...y pressing the Up or Down button followed by the E button to return to the alarm menu 8 3 7 Hysteresis HStr During programming hysteresis is shown in the units the indicator has been calibrated to dis...

Страница 17: ...security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to...

Страница 18: ...18 Fig 12 Alarm programme menu...

Страница 19: ...these functions the indicator initially displays the current break point and the total number of break points being used e g 5 13 current total number of break point break points Add Add a break point...

Страница 20: ...ak points Each subsequent operation of the P push button will introduce an additional break point When adding a break point to a calibrated indicator the insertion position can be selected using the U...

Страница 21: ...21...

Страница 22: ...tivated The tare function allows an operator to quickly zero the indicator display at any input current so that subsequent readings only show the change in input since the P push button was operated W...

Отзывы: