5.7 Conditioning sub-menu:  Cond

This sub-menu allows the mains (line) frequency at
which the indicator has maximum ac rejection to be
selected and the two internal references to be
calibrated. These functions are contained in a sub-
menu to minimise the possibility of inadvertent
adjustment. To gain access to the sub-menu select
'Cond' from the main menu and press 

P

. The

indicator will display 'SurE' to warn that changing
the parameters in the sub-menu will change the
indicator display. Pressing 

P

 again will give access

to the sub-menu, or pressing 

E

 will return the

indicator to the main menu.

5.7.1 AC rejection:  FrE

Caution

If the mains filter frequency is
changed, the indicator display and
internal references (if used) must
be recalibrated.

To provide maximum low frequency rejection the
internal digital filter may be set to operate at 50 or
60Hz to correspond with the local mains (line)
frequency. To change the frequency select 'FrE'
from the Cond sub-menu and press 

P

 which will

reveal the current setting. The setting can be
changed by pressing the 

Up

 or 

Down

 buttons

followed by the 

E

 button to return to the sub-menu.

5.7.2 Calibration of internal references:  rEF

The indicator contains two references representing
4 and 20mA. These references are used in the
'SEt' function which enables the indicator display to
be calibrated without an external current calibrator.
They are also used in the 'C--P' function when the

P

 push-button is programmed to display the input

current in the display mode. If neither of these
functions is to be used, it is not necessary to
calibrate the internal references.

To calibrate the references select 'rEF' from the
sub-menu and press 

P

 which will result in a

'0.004A' prompt being displayed. Adjust the
external current calibrator to 4.000mA and again
press 

P

. The indicator will display 'Ent' when the

4mA reference has been updated and will then
return to the '.004A' prompt.

To re-calibrate the 20mA internal reference, press
the 

Up

 button which will cause the indicator to

display '.020A'. Adjust the external current
calibrator to 20.000mA and again press 

P

. The

indicator will display 'Ent' when the 20mA
reference has been updated and will then return to
the '.020A' prompt. Two operations of the 

E

 button

will return the indicator to the main menu.

The accuracy of the internal references, and hence
the display accuracy, will depend upon the
accuracy of the external current source. With a
maximum span of 19999 the indicators have a
display resolution of 0.8µA, we therefore
recommend that the accuracy of the external
current source used for calibration is greater than
0.4µA.

5.8 Function of the P push-button:  C - - P

This parameter defines the function of the 

P

pushbutton when the indicator is in the display
mode. While the button is operated the indicator
will display the input current in milliamps, or the
input current as a percentage of the span.

To check or change the parameter select 'C - -P'
from the main menu and press 

P

 to reveal the

current setting. Pressing the 

Up

 or 

Down

 button

will toggle the setting between '4-20' the current
display and 'PC' the percentage display. When set
as required press 

E

 to return to the main menu.

Accuracy of the current display depends upon the
accuracy of the internal references which should
be periodically calibrated - see section 5.7.2

5.9 Security code:  COdE

The calibration and conditioning of the instrument
may be protected by a four digit security code
which must be entered before access to the
programme menu is granted. New instruments are
programmed with the default security code 0000
which allows unrestricted access to all
programming functions.

To enter a new security code select 'COdE' from
the menu and press 

P

 which will cause the

indicator to display the current security code. Each
digit of the code can be changed using the 

Up

 or

Down

 push-button, and the 

P

 button to move to the

next digit. When the required code has been
entered press 

E

 to return to the main menu. The

revised security code will be activated when the
indicator is returned to the operating mode.

If the security code is lost, access to the
programmable functions can be obtained by
moving the internal security link to the override
position. The original security code can then be
viewed by selecting 'CodE' from the main menu
and pressing 

P

.

Содержание BA527C

Страница 1: ...Issue 3 5th March 2009 BA527C BA528C loop powered 4 digit panel mounting indicators Issue 3...

Страница 2: ...urce 6 2 Using the internal calibrator 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comme...

Страница 3: ...o the 4 20mA input current provides the input signal for the analogue to digital converter Fig 1 Simplified block diagram Each time a 4 20mA current is applied to the instrument initialisation is perf...

Страница 4: ...any 4 20mA signal to provide a remote indication Fig 3 shows a typical application in which the output from a gas analyser drives a BA527C or BA528C indicator Again it is necessary to ensure that the...

Страница 5: ...trument and tighten until the indicator is secure as shown in Fig 5 Four clips are required to achieve an IP65 between a BA528C and the instrument panel Fig 4B BA528C dimensions c Connect the panel wi...

Страница 6: ...this manual push buttons are shown in italics e g P or Up push button and legends displayed by the indicator are shown within inverted commas e g CAL and ALr2 Access to the programme menu is obtained...

Страница 7: ...0mA loop See section 5 6 Display Description of function Cond Indicator conditioning This function provides access to a sub menu enabling the internal references to be calibrated and the mains line fr...

Страница 8: ...8...

Страница 9: ...adjusted press E to enter the new zero The indicator will display Ent for a few seconds while the information is being stored in memory and will then return to the ZErO prompt To adjust the display at...

Страница 10: ...alibrator to 20 000mA and again press P The indicator will display Ent when the 20mA reference has been updated and will then return to the 020A prompt Two operations of the E button will return the i...

Страница 11: ...requency of the mains supply is 50Hz and the existing security code is 1209 In the operating mode the indicator is required to display the input current as a percentage of span when the P push button...

Страница 12: ...ain menu Step 8 Return to the display mode Return to the display mode by pressing E 6 2 Using the internal calibrator As in 6 1 the BA527C is required to display 50 0 with a 4mA input 1050 0 with a 20...

Страница 13: ...ing If a BA527C or BA528C fails to function during commissioning the following procedure should be followed Symptom Cause Solution No display Incorrect There should be 1V wiring between terminals 1 3...

Страница 14: ...d 7 6 Customer comments BEKA associates is always pleased to receive comments from customers about our products and services All communications are acknowledged and whenever possible suggestions are i...

Страница 15: ...application of alarms Summary of programmable alarm functions Display Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section...

Страница 16: ...y pressing the Up or Down button followed by the E button to return to the alarm menu 8 3 7 Hysteresis HStr During programming hysteresis is shown in the units the indicator has been calibrated to dis...

Страница 17: ...security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to...

Страница 18: ...18 Fig 12 Alarm programme menu...

Страница 19: ...these functions the indicator initially displays the current break point and the total number of break points being used e g 5 13 current total number of break point break points Add Add a break point...

Страница 20: ...ak points Each subsequent operation of the P push button will introduce an additional break point When adding a break point to a calibrated indicator the insertion position can be selected using the U...

Страница 21: ...21...

Страница 22: ...tivated The tare function allows an operator to quickly zero the indicator display at any input current so that subsequent readings only show the change in input since the P push button was operated W...