
4.2 Using Galvanic Isolators
Galvanic isolators, although more expensive than
Zener barriers, do not require a high integrity earth
connection. For small systems where a high
integrity earth is not already available, the use of a
galvanic isolator often reduces overall installation
cost and will simplify the design of the loop.
Galvanic isolators also enable higher resistance
loads to be driven by the 4/20mA current.
Any EEx ia IIC certified isolator with output safety
parameters equal to or less than the following may
be used:
Uo
=
30V
Io
=
280mA
Po
=
0.85W
For guidance System Certificate Ex96D2506.
lists some of the devices which may be used
Fig 4 shows a typical loop.
Fig 4 Typical application using galvanic isolators
5. INSTALLATION
The BA378C indicating temperature transmitter
should only be installed and operated by trained
competent personnel.
5.1 Location
The BA378C indicating transmitter is housed in a
robust aluminium enclosure with a polyester front
panel and Noryl bezel. The front panel provides
IP65 protection and a gasket seals the joint
between the instrument and the panel. The trans-
mitter must be positioned such that the environ-
mental conditions specified by the instrument
datasheet and the EC-Type Examination Certificate
are not exceeded, and it will not be attacked by
aggresive substances.
Please consult BEKA associates if the instrument is
to be installed in an area subject to high vibration.
Fig 5 shows the overall dimensions of the BA378C.
Recommended panel cut-outs are as detailed
below.
a) DIN 43 700
138.0 +1.0 / -0.0mm
68.0
+0.7 / -0.0mm
b) To achieve IP65 seal between instrument and
panel
136.0 +0.5 / -0.0mm
66.2 +0.5 / -0.0
mm
Fig 5 BA378C dimensions
5.2 Installation Procedure
Figs 6a and 6b illustrate the instrument installation
procedure.
a.
Insert the instrument into the panel aperture
from the front of the panel as shown in Fig
6a.
b.
Fix two panel mounting clips to opposite sides
of the instrument and tighten until the
instrument is secure.
Four clips will be required to achieve an IP65
seal between the instrument and the panel.
c.
Connect the panel wiring to the rear terminal
block(s) as shown in Fig 6b.
WARNING!
To prevent damage to the static
dissipative coating on the front panel &
bezel of the instrument, only soap and
water should be used for cleaning
Solvents and abrasives must not be
used.
8