background image

25

Summary of programmable alarm

functions

Display

      Description of function

'EnbL'

Alarm enable

Enables or disables the  alarm
function without changing the
alarm parameters.
See section  9.3.4

'tYPE'

Type of alarm

Defines whether the alarm
operates on the rate or total
display.
See section 9.3.5

'SPr1'

Alarm setpoint 1

   or 

Adjusts  the  alarm  setpoint .  The

'SPt1'

alarm  is activated when the rate or
total display equals the setpoint.
See section 9.3.6

'HI.LO'

Alarm function

Defines whether the alarm has a
high or low function
See section 9.3.7

'no.nc'

Normally open or normally
closed output

Determines whether the single
pole alarm output is open or closed
in the non alarm condition.
See section 9.3.8

'HStr'

Hysteresis

Adjusts the alarm hysteresis.  Only
available on rate alarms.
See section 9.3.9

'dELA'

Alarm delay time

Adjusts the delay between the
display equalling the setpoint and
the alarm output being activated.
See section 9.3.10

'SIL'

Alarm silence time

Defines the time that the alarm
output remains in the non-alarm
condition following acceptance of
an alarm.
See section 9.3.11

'AcSP'

Access setpoint

Sub-menu which enables direct
access to the alarm setpoints from
the display mode and defines a
separate security code.
See section  9.3.12

9.3.4  Alarm enable:  EnbL

This function allows the alarm to be enabled
or disabled without altering any of the alarm
parameters.  To check or change the
function select 'EnbL' from the alarm menu
and press P which will reveal the current
setting.  The function can be changed by
pressing the Up or Down button followed by
the E button to return to the alarm menu.

9.3.5  Type of alarm:  tYPE

This function conditions the alarm to operate
on the rate or total display.   Alarm 1 and
Alarm 2 may both be conditioned as rate or
total alarms, or one may be conditioned for
rate and the other for total.

To check or change the type of alarm, select
'tYPE' from the alarm menu and press P
which will reveal the current setting.  The
function can be changed by pressing the Up
or  Down button followed by the E button to
return to the alarm menu.

Note:

  To prevent accidental setting, the

alarm must be re-enabled after the alarm
type has been changed  – see 9.2.4

9.3.6  Setpoint adjustment:
          SPx1 and SPx2

The rate alarm setpoints SPr1 and SPr2
may be positioned anywhere between  0 and
20000 providing this corresponds to an input
current between 3.8 and 20.2mA.   The total
alarm setpoint SPt1 and SPt2 may be
positioned anywhere between 00000000
and 99999999 To adjust the setpoint select
'AL1'  or ‘AL2’ from the alarm menu and
press  P which will reveal the existing alarm
setpoint.  Each digit of the setpoint can be
adjusted using the Up and Down push-
buttons, and the P button to transfer control
between digits.  When the required setpoint
has been entered press E  to return to the
alarm menu.

9.3.7  Alarm function:   HI.LO

Each alarm can be conditioned as a high or
low alarm.   To check or change the alarm
function select 'HI.LO' from the alarm menu
and press P to reveal the current setting.
The function can be changed by pressing
the  Up or Down buttons followed by the E
button to return to the alarm menu.

Содержание BA358C

Страница 1: ...October 2010 BA358C Intrinsically safe 4 20mA loop powered panel mounting rate totaliser Issue 4 Certification information label Rate display Annunciators for optional alarms Rotating flow indicator T...

Страница 2: ...1 Timebase 6 12 Function of the P push button 6 13 Total scale factor 6 14 Percentage clip off 6 15 Local reset of total display 6 16 Security code 6 17 Over and under range 7 Calibration Examples 7 1...

Страница 3: ...sed diode D1 The voltage developed across D1 which is relatively constant is multiplied by a switch mode power supply and used to power the instrument The voltage developed across R1 which is proporti...

Страница 4: ...eds 4mA 3 INTRINSIC SAFETY CERTIFICATION 3 1 ATEX certificate The BA358C has been issued with an EC Type Examination Certificate by Notified Body Intertek Testing Services ITS formerly ERA Technology...

Страница 5: ...Zones gas groups and T rating The BA358C has been certified to show compliance with the requirements for Group II Category 1G equipment EEx ia IIC T5 When connected to a suitable system the instrumen...

Страница 6: ...dication of rate and total flow within a hazardous area The type of interface is not critical either a Zener barrier or a galvanic isolator may be used providing that Uo Io and Po are not exceeded If...

Страница 7: ...esetting from the safe area via a single channel diode return Zener barrier A certified intrinsically safe relay or certified opto isolator may also be used Fig 3B shows how the total display may be r...

Страница 8: ...lective displays there is little degradation of contrast when viewed from above the centre line but slight degradation may be noticeable when the instrument is fitted with a backlight Fig 4 BA358C dim...

Страница 9: ...in the rate display is defined by the d p function To simplify calibration the BA358C uses floating point arithmetic moving the rate display decimal point position will therefore affect totalisation T...

Страница 10: ...of 0 00 with 4mA input and 100 00 with 20mA input The total scale factor will be set to 1 0000 with a timebase of flow per second The instrument can easily be recalibrated on site 6 3 Summary of progr...

Страница 11: ...input current in milliamps or the input as a percentage when the P push button is operated in the display mode Select PC for of span and 4 20 to display current See section 6 12 Display Description o...

Страница 12: ...cimal point clip off to adjust value Press or between 0 and 10 or and pressing Define Security Enter by Code to move to next digit Local reset or Fig 7B Programme Structure CLr Gtot E E E E E E E E P...

Страница 13: ...or and Similarly set the required full scale display accurate 20mA input current scale display using an Similarly set the required full to move to next digit pressing or and set required zero display...

Страница 14: ...s Operating the P push button will toggle control between the two displays When both decimal points have been correctly positioned press E to return to the main menu If the application requires flow r...

Страница 15: ...ay 0000 Using the Up and Down buttons and the P button to move to the next digit enter the password SurE S is entered as 5 Pressing E will then give access to the sub menu 6 9 1 AC rejection FrE WARNI...

Страница 16: ...our Tb 3600 3600 Select the required divisor and press E to return to the main menu 6 12 Function of the P push button C P This parameter defines the function of the P push button when the BA358C is i...

Страница 17: ...lect CLIP OFF from the main menu and press P to reveal the current setting This clip off threshold is shown as a percentage of total flow and may be adjusted in one percent increments using the Up or...

Страница 18: ...ction 7 1 describes the preferred method of calibration using an external traceable 4 20mA calibrator Section 7 2 contains an example of calibration using the BA358C internal 4 and 20mA references whi...

Страница 19: ...git of the total display flashing Press the Up or Down push button until the second least significant digit and following decimal point of the total display flash This gives the required total display...

Страница 20: ...to the main menu by pressing E Step 12 Reset the grand total to zero To reset the grand total to zero select CLr Gtot from the main menu and press P which will cause the instrument to display Clr no T...

Страница 21: ...multaneously pressing P and E Assuming a security code has not already been entered the instrument will respond by displaying root which is the first function in the main menu Steps Perform as describ...

Страница 22: ...terminals 1 3 No power supply Confirm that a 4 20mA current is flowing in the loop Unstable rate display 4 20mA current is noisy or the flow signal is unstable Eliminate source of electrical noise Red...

Страница 23: ...ons available for a rate alarm and shows which are fail safe i e output is in the alarm condition open when the 4 20mA input current is zero Similarly when programmed as a total alarm the output shoul...

Страница 24: ...nt capacitance and inductance of each BA358C alarm output is Ci 40nF Li 20 H The maximum permissible cable parameters for the alarm circuit will be the figures specified for the barrier or isolator le...

Страница 25: ...veal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return to the alarm menu 9 3 5 Type of alarm tYPE This function conditions the alarm...

Страница 26: ...Down button followed by the E button to return to the alarm menu 9 3 9 Hysteresis HStr Hysteresis is only available on rate alarms During programming hysteresis is shown in the units of rate the BA358...

Страница 27: ...This direct access menu is enabled and a separate security code entered from the AcSP function in the main programme menu To change the menu parameters select AcSP from the programme menu and press P...

Страница 28: ...is segregated from the measuring circuit and has been certified as a separate intrinsically safe circuit but it does not comply with the requirements for simple apparatus Systems certificates for use...

Страница 29: ...ere the instrument is likely to be damaged by impact an additional front cover is available The BA398 is a robust transparent cover which is secured to the panel by six screws and sealed with a neopre...

Страница 30: ...missioning 8 1 after commissioning 8 2 Flow indicator 2 2 Flow transmitter loops 4 1 Galvanic isolators 4 0 Gas groups 3 4 Grand total 2 1 clear 6 3 6 10 Guarantee 8 5 HI LO 9 3 3 9 3 7 HStr 9 3 3 9 3...

Отзывы: