background image

16

viewed in the operating mode in two eight
digit sections as described in section 2.1 of
this manual.

The grand total counter can only be reset to
zero from the 'CLr.Gtot' function within the
main programme menu.   To zero the grand
total counter select 'CLr. Gtot' and press P
which will cause the instrument to display
'Clr. no' with 'no' flashing.    Press the Up or
Down button until 'CLr. YES' is displayed
followed by P which will result in a '0000'
prompt.  Using the Up and Down buttons
and the P button to move to the next digit,
confirm the request by entering the
password 'SurE'.  Note 'S' is entered as '5'.
Pressing  E will then reset the grand total
counter to zero and return the instrument to
the 'CLr. Gtot' prompt in the main menu.

WARNING

After resetting the grand total counter
to zero the old grand total can not be
recovered.

6.11  Timebase:  t-bASE

Together with the total scale factor 'SCALE-
t', this function defines the arithmetic
relationship between the rate and total
displays.    The timebase divides the rate
display by 1, 60 or 3,600 depending upon
whether the BA358C rate display represents
flow per second, per minute or per hour.
See Fig 6.

To check  or  change the timebase select
't-bASE' from the main menu and press P
which will reveal the current setting.
Pressing the Up or Down buttons will index
though the following three options.

BA358C

Rate

display

represents

flow

Select

In

t-base

menu

Total display

= rate

divided by

Per second

tb – 1

1

Per minute

Tb – 60

60

Per hour

Tb – 3600

3600

Select the required divisor and press E to
return to the main menu.

6.12  Function of the 

P

 push-button:

         C - - P

This parameter defines the function of the P
push-button when the BA358C is in the
display mode.   Either the input current in
milliamps or the input as a percentage of
span may be shown.

To check or change this parameter select 'C
- -P' from the main menu and press P to
reveal the current setting.  Pressing the Up
or  Down button will toggle the instrument
between the two options.  Select '4-20' if the
input current in milliamps is required, or 'PC'
if the input as a percentage of span is
required.  When the required option is
displayed press E to store the selection and
return to the main menu.

Accuracy of both these options depends
upon the accuracy of the internal references
which should be periodically calibrated - see
section 6.9.2

6.13  Total scale factor:  SCALE-t

Together with the timebase 't-bASE', this is a
dividing factor which defines the arithmetic
relationship between the rate and total
displays.  The total scale factor may be set
to any value between 0.0001 and 65535*
allowing the total flow and rate of flow to be
displayed in different engineering units.

SCALE-t  =    Units of rate display

  

                   Units of total display

When calculating SCALE-t the required units
of measurement for the total display must be
converted to those used for the rate display
as shown in following example:

The BA358C rate display represents litres
per second but the total flow display is
required in UK gallons.   There are 4.5461
litres in a UK gallon.

SCALE-t  =    Units of rate display
                      Units of total display

      =    1 litre     =         1 litre

  

         1 gallon     1/(4.5461) litres

SCALE-t  =    4.5461
  

Содержание BA358C

Страница 1: ...October 2010 BA358C Intrinsically safe 4 20mA loop powered panel mounting rate totaliser Issue 4 Certification information label Rate display Annunciators for optional alarms Rotating flow indicator T...

Страница 2: ...1 Timebase 6 12 Function of the P push button 6 13 Total scale factor 6 14 Percentage clip off 6 15 Local reset of total display 6 16 Security code 6 17 Over and under range 7 Calibration Examples 7 1...

Страница 3: ...sed diode D1 The voltage developed across D1 which is relatively constant is multiplied by a switch mode power supply and used to power the instrument The voltage developed across R1 which is proporti...

Страница 4: ...eds 4mA 3 INTRINSIC SAFETY CERTIFICATION 3 1 ATEX certificate The BA358C has been issued with an EC Type Examination Certificate by Notified Body Intertek Testing Services ITS formerly ERA Technology...

Страница 5: ...Zones gas groups and T rating The BA358C has been certified to show compliance with the requirements for Group II Category 1G equipment EEx ia IIC T5 When connected to a suitable system the instrumen...

Страница 6: ...dication of rate and total flow within a hazardous area The type of interface is not critical either a Zener barrier or a galvanic isolator may be used providing that Uo Io and Po are not exceeded If...

Страница 7: ...esetting from the safe area via a single channel diode return Zener barrier A certified intrinsically safe relay or certified opto isolator may also be used Fig 3B shows how the total display may be r...

Страница 8: ...lective displays there is little degradation of contrast when viewed from above the centre line but slight degradation may be noticeable when the instrument is fitted with a backlight Fig 4 BA358C dim...

Страница 9: ...in the rate display is defined by the d p function To simplify calibration the BA358C uses floating point arithmetic moving the rate display decimal point position will therefore affect totalisation T...

Страница 10: ...of 0 00 with 4mA input and 100 00 with 20mA input The total scale factor will be set to 1 0000 with a timebase of flow per second The instrument can easily be recalibrated on site 6 3 Summary of progr...

Страница 11: ...input current in milliamps or the input as a percentage when the P push button is operated in the display mode Select PC for of span and 4 20 to display current See section 6 12 Display Description o...

Страница 12: ...cimal point clip off to adjust value Press or between 0 and 10 or and pressing Define Security Enter by Code to move to next digit Local reset or Fig 7B Programme Structure CLr Gtot E E E E E E E E P...

Страница 13: ...or and Similarly set the required full scale display accurate 20mA input current scale display using an Similarly set the required full to move to next digit pressing or and set required zero display...

Страница 14: ...s Operating the P push button will toggle control between the two displays When both decimal points have been correctly positioned press E to return to the main menu If the application requires flow r...

Страница 15: ...ay 0000 Using the Up and Down buttons and the P button to move to the next digit enter the password SurE S is entered as 5 Pressing E will then give access to the sub menu 6 9 1 AC rejection FrE WARNI...

Страница 16: ...our Tb 3600 3600 Select the required divisor and press E to return to the main menu 6 12 Function of the P push button C P This parameter defines the function of the P push button when the BA358C is i...

Страница 17: ...lect CLIP OFF from the main menu and press P to reveal the current setting This clip off threshold is shown as a percentage of total flow and may be adjusted in one percent increments using the Up or...

Страница 18: ...ction 7 1 describes the preferred method of calibration using an external traceable 4 20mA calibrator Section 7 2 contains an example of calibration using the BA358C internal 4 and 20mA references whi...

Страница 19: ...git of the total display flashing Press the Up or Down push button until the second least significant digit and following decimal point of the total display flash This gives the required total display...

Страница 20: ...to the main menu by pressing E Step 12 Reset the grand total to zero To reset the grand total to zero select CLr Gtot from the main menu and press P which will cause the instrument to display Clr no T...

Страница 21: ...multaneously pressing P and E Assuming a security code has not already been entered the instrument will respond by displaying root which is the first function in the main menu Steps Perform as describ...

Страница 22: ...terminals 1 3 No power supply Confirm that a 4 20mA current is flowing in the loop Unstable rate display 4 20mA current is noisy or the flow signal is unstable Eliminate source of electrical noise Red...

Страница 23: ...ons available for a rate alarm and shows which are fail safe i e output is in the alarm condition open when the 4 20mA input current is zero Similarly when programmed as a total alarm the output shoul...

Страница 24: ...nt capacitance and inductance of each BA358C alarm output is Ci 40nF Li 20 H The maximum permissible cable parameters for the alarm circuit will be the figures specified for the barrier or isolator le...

Страница 25: ...veal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return to the alarm menu 9 3 5 Type of alarm tYPE This function conditions the alarm...

Страница 26: ...Down button followed by the E button to return to the alarm menu 9 3 9 Hysteresis HStr Hysteresis is only available on rate alarms During programming hysteresis is shown in the units of rate the BA358...

Страница 27: ...This direct access menu is enabled and a separate security code entered from the AcSP function in the main programme menu To change the menu parameters select AcSP from the programme menu and press P...

Страница 28: ...is segregated from the measuring circuit and has been certified as a separate intrinsically safe circuit but it does not comply with the requirements for simple apparatus Systems certificates for use...

Страница 29: ...ere the instrument is likely to be damaged by impact an additional front cover is available The BA398 is a robust transparent cover which is secured to the panel by six screws and sealed with a neopre...

Страница 30: ...missioning 8 1 after commissioning 8 2 Flow indicator 2 2 Flow transmitter loops 4 1 Galvanic isolators 4 0 Gas groups 3 4 Grand total 2 1 clear 6 3 6 10 Guarantee 8 5 HI LO 9 3 3 9 3 7 HStr 9 3 3 9 3...

Отзывы: