background image

21

Step 6

Enter the rate timebase

Select ‘t-bASE’ from the main menu
and press P to show the current setting.
In this example the rate display is
required in litres per hour. Using the Up
or Down button scroll through the three
options and select ‘tb-3600’ which will
multiply the rate display by 3600.
Return to the ‘t-bASE’ prompt in the
main menu by pressing E.

Step 7

Adjust the rate filter

The rate display filter parameters
should be adjusted experimentally after
installation to provide a stable rate
display with an acceptable step
response.

During commissioning it is recommend
that initially the second digit of the rate
parameters is set to 0 (step response
off) and the first digit (amount of
filtering) is adjusted to provide
acceptable rate display stability. The
second digit should them be increased
until acceptable rate display stability is
once again achieved.

To adjust the filter parameters select
‘FILtEr’ from the main menu and press
P to reveal the current setting.  The first
digit will be flashing and may be
adjusted using the Up or Down  button.
The P button will transfer control to the
second digit.  When both are set as
required, return to the ‘FILtEr’ prompt in
the main menu by pressing E.

   Note:

  While adjusting the filter, the rate is

shown on the lower display so that
stability can be assessed.

Step 8    Enter the total scaling factor

The K factor of the flowmeter in this
example is 1050.0 pulses per litre and
the total display is required in cubic
metres.

There are 1000 litres in a cubic metre
so the BA338C is required to display 1
cubic metre for every 1,050 x 1,000
pulses received.  ‘SCALE-t’ should
therefore be set to 1,050,000

Select ‘SCALE-t’ from the main menu

and press P which will reveal the
current setting with one digit flashing.
To position the decimal point, operate
the P push-button to move the flashing
digit to the least significant position.
Pressing the Up and Down buttons
simultaneously will then remove the
decimal point.

Using the Up and Down buttons to
adjust each digit in turn and the P
button to transfer control between
digits, enter 1050000    Finally return to
the 'SCALE-t' prompt in the main menu
by pressing E.

Step 9

Define clip-off

In this example totalisation is to be
inhibited at flow rates below 300 litres
per hour.

Select ‘CLIP-OFF’ from the main menu
and press P which will reveal the clip-
off threshold in litres per hour i.e. the
same units already selected for the rate
display.  The most significant digit of
the display will be flashing, indicating
that this digit may be adjusted.
Pressing  P  five times will move the
flashing digit to the third lease
significant digit.

Using the Up and Down buttons set the
third least significant digit to 3  i.e. a
clip-off threshold of 300.     Finally
return to the 'CLIP-OFF' prompt in the
main menu by pressing E.

.

Step 10 Turn local reset on

In this example the operator needs to
reset the total display to zero by
pressing the Up and Down buttons
simultaneously.  To activate the local
reset function select ‘LOC.rSEt’ from
the main menu and press P to display
the current setting.  Pressing the Up or
Down button will toggle the display
between ‘On’ and ‘OFF’.  Select ‘On’
and press E to return to the ‘LOCrSEt’
prompt in the main menu.

Содержание BA338C

Страница 1: ...ry 2011 Certification information label Rate display Annunciators for optional alarms Rotating flow indicator Total display BA338C Intrinsically safe Externally powered pulse input panel mounting rate...

Страница 2: ...te interval 6 5 Type of input 6 6 Position of the decimal points 6 7 Lower display 6 8 Rate scale factor 6 9 Timebase 6 10 Rate filter 6 11 Total scale factor 6 12 Clip off 6 13 Local reset 6 14 Reset...

Страница 3: ...ost any pulse output flowmeter The total display may be reset to zero by a remote switch contact or from the front panel push buttons Optional accessories enable the flow rate to be transmitted to oth...

Страница 4: ...fied Body Intertek Testing Services ITS confirming compliance with the European ATEX Directive 94 9 EC for Group II Category 1 gas atmospheres EEx ia IIC T5 The instrument bears the Community Mark and...

Страница 5: ...lvanic isolator connected to terminals 3 and 4 The two system certificates specify maximum permitted cable parameters when 28V 300 devices are used 3 3 2 Contact 2 wire proximity detector or open coll...

Страница 6: ...nfirming that it complies with the requirements for Group II Category 1 G EEx ia IIC T5 Tamb 40 to 60o C specified in the ATEX Directive When connected to a suitable system the BA338C may be installed...

Страница 7: ...uit that is used for all BA338C installations protected by Zener barriers For simplicity connections for the optional pulse output 4 20mA output and backlight are included in the appendix to this manu...

Страница 8: ...tified intrinsically safe voltage source within the hazardous area which can withstand a 500V rms insulation test to earth for one minute and has output parameters equal to or less than Uo 28V dc Io 1...

Страница 9: ...is used to transfer the signal to the rate totaliser in the hazardous area Fig 5 Pulse source in safe area 4 3 1 Power supply Any certified galvanic isolator may be used to power the BA338C providing...

Страница 10: ...ces should be connected to terminals 3 and 4 These terminals comply with the requirements for simple apparatus They may be directly connected to any certified intrinsically safe voltage source within...

Страница 11: ...ent secured with four panel mounting clips The BA338C liquid crystal display has maximum contrast when viewed from directly ahead and slightly below the centre line of the instrument For reflective di...

Страница 12: ...ith the requirements of the European EMC Directive 2004 108 EEC For specified immunity all wiring should be in screened twisted pairs To prevent circulating currents cable screens should only be earth...

Страница 13: ...decimal points Moving the position of a decimal point in a scale factor will therefore affect the instrument calibration The BA338C updates its total flow calculation twice per second If the instrumen...

Страница 14: ...14...

Страница 15: ...AY 2 Lower display Turns the loweer display which normally shows rate on or off See section 6 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relationship between th...

Страница 16: ...sing Up or Down will scroll through the five options UOLts Voltage pulse input Threshold 1V and 3V 5kHz max Input terminals 3 and 4 COIL Low voltage pulse Threshold 20mV peak 5kHz max Input terminals...

Страница 17: ...on the decimal point in the rate scale factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and Down buttons When the required ra...

Страница 18: ...next digit To position the decimal point in the scaling factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and Down buttons Wh...

Страница 19: ...6 15 Security code CodE The calibration and conditioning of the instrument may be protected by a four digit security code which must be entered before access to the programme menu is granted New inst...

Страница 20: ...nput Using the Up or Down button scroll through the main menu until InPut is displayed then press P to reveal the existing setting Select COIL the input for a magnetic pick off using the Up or Down bu...

Страница 21: ...c metre so the BA338C is required to display 1 cubic metre for every 1 050 x 1 000 pulses received SCALE t should therefore be set to 1 050 000 Select SCALE t from the main menu and press P which will...

Страница 22: ...trument to the CLr Gtot prompt in the main menu Step 12 Define the security code Defining a security code prevents unauthorised access to the programme functions Select COdE from the main menu and pre...

Страница 23: ...alarms 8 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING MAINTENANCE Live maintenance is permitted on intrinsically safe equipment installed in a hazardous area but only certif...

Страница 24: ...status of both alarms If an alarm delay or silence time has been selected the annunciator will flash during the delay or silence period Programmable functions for each alarm include adjustable setpoin...

Страница 25: ...ay Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 9 3 4 tYPE Type of alarm Defines whether the alarm operates on the...

Страница 26: ...or the alarm type has been changed Both alarms must therefore be enabled after calibration changes 9 3 5 Type of alarm tYPE Alarm 1 and Alarm 2 may both be rate or total alarms or one may be conditio...

Страница 27: ...resis select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next digit When th...

Страница 28: ...l disable the security code allowing direct access to the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code...

Страница 29: ...ght powered by Zener barrier The display brilliance depends upon the current flowing through the backlight This is determined by the supply voltage and the end to end resistance of the Zener barrier o...

Страница 30: ...ce and inductance specified by the system certificate of the circuit connected to the BA338C Fig 16 shows how the pulse output circuit may be powered from the BA338C supply to drive a counter in the s...

Страница 31: ...the rate display by the d P function The value of the flashing digit may be changed by pressing the Up or Down buttons When this digit is set as required pressing P will transfer control to the next...

Страница 32: ...4 Fault finding during commissioning 8 1 after commissioning 8 2 FILtEr 6 10 Flow indicator 2 3 FM Approval Appendix 1 Galvanic isolators 4 3 Gas groups 3 5 Grand total 2 2 clear 6 14 Guarantee 8 5 HI...

Страница 33: ...d with the following gases Intrinsic Safety Division 1 or 2 Class I Group A B Group C Group D Zone 0 1 or 2 Class 1 Group IIC Group IIB Group IIA The FM entity parameters are similar to the ATEX param...

Страница 34: ...34...

Страница 35: ...35...

Страница 36: ...36...

Страница 37: ...37...

Страница 38: ...38...

Страница 39: ...39...

Страница 40: ...40...

Страница 41: ...41...

Отзывы: