background image

20

7.  CALIBRATION EXAMPLE

In this example a BA338C rate totaliser is
connected to a turbine flowmeter having a K-factor
of 1050.0 pulses per litre.   The flowmeter has a
magnetic pick-off with an  output greater than
20mV peak at 5 litres per minute and a usable
range of 5 to 60 litres per minute.

The BA338C is required to display rate of flow in
litres per hour with a resolution of one litre and
total flow in cubic metres with a resolution of 0.01
cubic metres.   Totalisation is to stop when the flow
rate falls below 300 litres per hour.  The display is
to be updated twice per second and filtering is
required.   In this application the operator needs to
reset the total display to zero by simultaneously
pushing the Up and Down push-buttons.  To
prevent tampering the instrument programme
menu is to be protected by security code 1209

7.1 Calibration procedure

The BA338C rate totaliser may be calibrated on-
site without disconnection from the power supply
or from the flowmeter.

Step 1

Enter the programming mode

Put the BA338C in the programming
mode by simultaneously pressing  P
and E.  Assuming a security code has
not already been entered the
instrument will respond by displaying
'UPdAtE' which is the first function in
the main menu.  See Fig 9.

Step 2

Select the interval between display
updates

With 'UPdAtE' displayed, press P    to
reveal the existing interval between
display updates.  If this is not as
required, press the Up or Down button
until ‘0.5’ is displayed. (0.5 seconds i.e.
2 display updates per second).  Enter
the revised time and return to the
'UPdAtE' prompt in the main menu by
pressing E.

Step 3

Select the type of input

Using the Up or Down button scroll
through the main menu until ‘InPut’ is
displayed, then press P to reveal the
existing setting.  Select  ‘COIL’, the
input for a magnetic pick-off, using the
Up or Down button and return to the
'InPut' prompt in the main menu by
pressing E.

Note:

 The BA338C has two pairs of

input terminals. 3 & 4 for voltage inputs
and 5 & 6 for switch contact, proximity
detector or open collector inputs.
Ensure that the magnetic pick-off that
has a voltage output is connected to
terminals 3 & 4.

Step 4

Position rate & total decimal points

Select ‘d.P.’ from the main menu and
press P.  The rate and total displays will
be activated with one digit of the total
display flashing.  Press the Up or Down
push-button until the third least
significant digit of the total display and
following decimal point are flashing.
This gives the required total display
resolution of 0.01

Press  P to move control to the rate
display. Using the Up or Down push-
button move the flashing digit to the
least significant position (right hand
side) which will result in no decimal
point being displayed.

Finally press E to return to the 'd.P’.
prompt in the main menu.

Step 5

Enter the rate scaling factor

Select ‘SCALE-r’ from the main menu
and press P to show the current figure.
The K factor of the flowmeter in this
example is 1050.0 pulses per litre
which should be entered as the rate
scaling factor.

Firstly to position the decimal point,
operate the P push-button to move the
flashing digit to second least significant
position.  Pressing the Up and Down
buttons simultaneously will then
position the decimal point in front of the
least significant digit.

Using the Up and Down buttons to
adjust each digit in turn and the P
button to transfer control between
digits, enter 1050.0     Finally return to
the 'SCALE-r' prompt in the main menu
by pressing E.

Содержание BA338C

Страница 1: ...ry 2011 Certification information label Rate display Annunciators for optional alarms Rotating flow indicator Total display BA338C Intrinsically safe Externally powered pulse input panel mounting rate...

Страница 2: ...te interval 6 5 Type of input 6 6 Position of the decimal points 6 7 Lower display 6 8 Rate scale factor 6 9 Timebase 6 10 Rate filter 6 11 Total scale factor 6 12 Clip off 6 13 Local reset 6 14 Reset...

Страница 3: ...ost any pulse output flowmeter The total display may be reset to zero by a remote switch contact or from the front panel push buttons Optional accessories enable the flow rate to be transmitted to oth...

Страница 4: ...fied Body Intertek Testing Services ITS confirming compliance with the European ATEX Directive 94 9 EC for Group II Category 1 gas atmospheres EEx ia IIC T5 The instrument bears the Community Mark and...

Страница 5: ...lvanic isolator connected to terminals 3 and 4 The two system certificates specify maximum permitted cable parameters when 28V 300 devices are used 3 3 2 Contact 2 wire proximity detector or open coll...

Страница 6: ...nfirming that it complies with the requirements for Group II Category 1 G EEx ia IIC T5 Tamb 40 to 60o C specified in the ATEX Directive When connected to a suitable system the BA338C may be installed...

Страница 7: ...uit that is used for all BA338C installations protected by Zener barriers For simplicity connections for the optional pulse output 4 20mA output and backlight are included in the appendix to this manu...

Страница 8: ...tified intrinsically safe voltage source within the hazardous area which can withstand a 500V rms insulation test to earth for one minute and has output parameters equal to or less than Uo 28V dc Io 1...

Страница 9: ...is used to transfer the signal to the rate totaliser in the hazardous area Fig 5 Pulse source in safe area 4 3 1 Power supply Any certified galvanic isolator may be used to power the BA338C providing...

Страница 10: ...ces should be connected to terminals 3 and 4 These terminals comply with the requirements for simple apparatus They may be directly connected to any certified intrinsically safe voltage source within...

Страница 11: ...ent secured with four panel mounting clips The BA338C liquid crystal display has maximum contrast when viewed from directly ahead and slightly below the centre line of the instrument For reflective di...

Страница 12: ...ith the requirements of the European EMC Directive 2004 108 EEC For specified immunity all wiring should be in screened twisted pairs To prevent circulating currents cable screens should only be earth...

Страница 13: ...decimal points Moving the position of a decimal point in a scale factor will therefore affect the instrument calibration The BA338C updates its total flow calculation twice per second If the instrumen...

Страница 14: ...14...

Страница 15: ...AY 2 Lower display Turns the loweer display which normally shows rate on or off See section 6 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relationship between th...

Страница 16: ...sing Up or Down will scroll through the five options UOLts Voltage pulse input Threshold 1V and 3V 5kHz max Input terminals 3 and 4 COIL Low voltage pulse Threshold 20mV peak 5kHz max Input terminals...

Страница 17: ...on the decimal point in the rate scale factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and Down buttons When the required ra...

Страница 18: ...next digit To position the decimal point in the scaling factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and Down buttons Wh...

Страница 19: ...6 15 Security code CodE The calibration and conditioning of the instrument may be protected by a four digit security code which must be entered before access to the programme menu is granted New inst...

Страница 20: ...nput Using the Up or Down button scroll through the main menu until InPut is displayed then press P to reveal the existing setting Select COIL the input for a magnetic pick off using the Up or Down bu...

Страница 21: ...c metre so the BA338C is required to display 1 cubic metre for every 1 050 x 1 000 pulses received SCALE t should therefore be set to 1 050 000 Select SCALE t from the main menu and press P which will...

Страница 22: ...trument to the CLr Gtot prompt in the main menu Step 12 Define the security code Defining a security code prevents unauthorised access to the programme functions Select COdE from the main menu and pre...

Страница 23: ...alarms 8 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING MAINTENANCE Live maintenance is permitted on intrinsically safe equipment installed in a hazardous area but only certif...

Страница 24: ...status of both alarms If an alarm delay or silence time has been selected the annunciator will flash during the delay or silence period Programmable functions for each alarm include adjustable setpoin...

Страница 25: ...ay Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 9 3 4 tYPE Type of alarm Defines whether the alarm operates on the...

Страница 26: ...or the alarm type has been changed Both alarms must therefore be enabled after calibration changes 9 3 5 Type of alarm tYPE Alarm 1 and Alarm 2 may both be rate or total alarms or one may be conditio...

Страница 27: ...resis select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next digit When th...

Страница 28: ...l disable the security code allowing direct access to the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code...

Страница 29: ...ght powered by Zener barrier The display brilliance depends upon the current flowing through the backlight This is determined by the supply voltage and the end to end resistance of the Zener barrier o...

Страница 30: ...ce and inductance specified by the system certificate of the circuit connected to the BA338C Fig 16 shows how the pulse output circuit may be powered from the BA338C supply to drive a counter in the s...

Страница 31: ...the rate display by the d P function The value of the flashing digit may be changed by pressing the Up or Down buttons When this digit is set as required pressing P will transfer control to the next...

Страница 32: ...4 Fault finding during commissioning 8 1 after commissioning 8 2 FILtEr 6 10 Flow indicator 2 3 FM Approval Appendix 1 Galvanic isolators 4 3 Gas groups 3 5 Grand total 2 2 clear 6 14 Guarantee 8 5 HI...

Страница 33: ...d with the following gases Intrinsic Safety Division 1 or 2 Class I Group A B Group C Group D Zone 0 1 or 2 Class 1 Group IIC Group IIB Group IIA The FM entity parameters are similar to the ATEX param...

Страница 34: ...34...

Страница 35: ...35...

Страница 36: ...36...

Страница 37: ...37...

Страница 38: ...38...

Страница 39: ...39...

Страница 40: ...40...

Страница 41: ...41...

Отзывы: