CAUTION

When adding break-points to a new indicator,
or following resetting of the lineariser to the
factory  defaults  using  the  'LtAb'  function
described  in  section  6.11,  new  break-points
should  be  added  between  the  two  default
break-points  '0:1'  and  '1:1'.      If  new  break-
points are added before breakpoint '0:1' which
has a default setting of 4mA, it will result in a
'FaiL'  error  message  when  the  indicator
display is entered. 

The delete break-point sub-function 'dEL' operates
in exactly the same way as the 'Add' sub-function
described  above.      Once  within  the  ‘dEL’  sub-
function each time the 

P

 button is pressed a break-

point is removed. When deleting a break-point from
a calibrated indicator, the break-point to be deleted
can be selected using the 

 and 

 push buttons.

The minimum number of break-point is 2, break-
points '0:1' and '1:1'.  

When  the  required  number  of  linearising  break-
points has been entered, return to the linearisation
sub-menu by pressing 

E

.  

The indicator will display

the 'Add' or 'dEL' prompt depending upon the last
function used.    Using the 'Pts' sub-function the
input current at which each break-point occurs and
the  corresponding  indicator  display  may  now  be
defined. 

Using the 

 or 

 button select the  'PtS' function

in the sub-menu and press 

P

  to enter the function

which will display the first break-point '0 : n', where
n  is  the  total  number  of  linearising  break-points
entered   –   see   Fig   9.     The   selected   linearising
break-point can be changed using the 

 and 

buttons.  When the required linearising break-point
has been selected set the indicator input current to
the exact value at which the break-point is required
and press  

P

*

.   Using the 

 and 

 buttons and

the  

P

  button  to  move  between  digits,  enter  the

required indicator display at this break-point.  

When set as required, press the 

E

 

push button

 

to

enter the required indicator display and return to
the sub-menu from which the next beak-point can
be selected.   

*  

Note:

 

The  indicator  input  current  must  be

adjusted to the required value before the 

button

is operated to enter the required indicator display.

Error message

If during calibration the indicator displays a
'FAiL' error message the current at which the
selected break-point is being set is not above
the proceeding break-point or is not below the
following break-point.   

When  all  the  break-points  have  been  calibrated
pressing  

E

 

twice  will  return  the  indicator  to  the

‘CAL’ function in the configuration menu.

7.2  Lineariser calibration using the internal 
       reference.

The  ‘SEt’  function  enables  the  lineariser  to  be
calibrated  without  the  need  for  an  accurate
external current source.  Throughout the calibration
the  indicator  input  current  may  be  any  value
between 4 and 20mA.    

The ‘SEt’ functions contains four sub-functions. 

Display

      Description of function

'Add'

Add a break-point

Adds  a  new  break-point  before  the
displayed break-point.  The calibration
of existing break-points is not changed,
but  the  identification  number  of  all
subsequent break-point is increased by
one.

'dEL'

Remove a break-point

Removes  the  displayed  break-point
and joins the preceding segment to the
following  segment with  a  straight  line.
The  identification  number  of  all
subsequent  break-points  is  decreased
by one.

'in' 

Defines the current at which break-
point occurs.

Enables  the  required  current  at  each
break-point  to  be  defined  without
having  to  input  an  accurate  input
current to the indicator.

'diSP'

Defines  indicator  display  at  break-
point.

Enables  the  indicator  display  at  each
break-point to be defined.

The number of break-point required should first be
entered using the 'Add' and 'dEL'  sub-functions.
In  both  these  sub-functions  the  indicator  initially
displays  the  current  break-point  and  the  total
number of break-point being used as shown below.

To   add   a   break-point   using   the  

 or  

 button

select 'SEt' from the configuration menu and press

P

 

which will result in the

 

'Add' sub-function prompt

being displayed. To enter the sub-function press 

P

which will reveal the current break-point and the
total  number  of  break-point  which  have  already
been entered.    Each subsequent operation of the

P

   push button will introduce an additional break-

Содержание BA307NE

Страница 1: ...Issue 3 1st Novemeber 2017 BA307NE BA327NE Ex nA Ex tc loop powered panel mounting indicators Issue 3...

Страница 2: ...figuration 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 Guarantee 8 6 Customer comments 9 Accessories 9 1 Scale...

Страница 3: ...European ATEX Directive 94 9 EC The certificate number has an X suffix indicating that the two indicators are subject to special condition for safe use For installations in Zone 2 the panel enclosure...

Страница 4: ...both instruments with a common Type Examination Certificate number ITS14ATEX48028X This has been used to confirm compliance with the European ATEX Directive for Group II Category 3GD equipment The in...

Страница 5: ...roups Group A propane Group B ethylene Group C hydrogen In gases that may safely be used with equipment having a temperature classification of T1 450o C T2 300o C T3 200o C T4 135o C T5 100o C At ambi...

Страница 6: ...ea should be individually fused and contain a means of isolation These two requirements are frequently satisfied by using switch fuse holders with easily removable fuses which can be extracted to achi...

Страница 7: ...is a satisfactory method at the low voltages and currents common in instrumentation systems Clear identification of and easy access to the means of isolation is essential for their effective use It i...

Страница 8: ...el enclosure cut out dimensions Fig 4 Dimensions 5 2 Installation Procedure a Cut the aperture specified in Fig 4 in the panel enclosure Ensure that the edges of aperture are de burred b Inspect the i...

Страница 9: ...ard showing the requested units of measurement if this information is not supplied when the indicator is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common uni...

Страница 10: ...10...

Страница 11: ...uration BA307NE BA327NE Access code CodE 0000 0000 Function FunC Linear Linear Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00...

Страница 12: ...in 16 segment adjustable lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or pr...

Страница 13: ...tween the two currents is greater than 4mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the i...

Страница 14: ...rompt in the configuration menu 6 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit security code which must be entered to gain access New instruments...

Страница 15: ...e This method allows direct calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this me...

Страница 16: ...16...

Страница 17: ...to the required value before the P button is operated to enter the required indicator display Error message If during calibration the indicator displays a FAiL error message the current at which the s...

Страница 18: ...oints has been defined and then return to the in sub function by pressing the E button The corresponding indicator display at each of the break points can now be defined using the diSP sub function Us...

Страница 19: ...rior of the instrument with dust or moisture Some end users may prefer not to permit live maintenance to minimise risk ENSURE PLANT SAFETY BEFORE STARTING MAINTENANCE Live maintenance within the hazar...

Страница 20: ...the enclosure panel being distorted or creeping when the indicator panel clamps are tightened and ensures that the joint between the indicator and the panel enclosure in which it is mounted remains se...

Страница 21: ...ed in Zone 2 is displaying the output from an Flameproof Ex d 2 wire transmitter located in Zone 1 Alarm 1 is switching an Ex e sounder in Zone 1 and alarm 2 is switching a sounder located in the safe...

Страница 22: ...22...

Страница 23: ...ndicator display equals the setpoint See section 9 4 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 4 6 no nC Normally open or normally closed output Sets the alarm out...

Страница 24: ...r fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Страница 25: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Страница 26: ...int plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 16 Fig 16 Displayed value and setpoints on bargra...

Страница 27: ...o or greater than 9V At lower supply voltages the backlight will continue to function but with reduced brilliance and will draw less current The ATEX certificate therefore only specifies the maximum s...

Страница 28: ...re four design requirements 1 The indicator should be installed generally as shown in section 5 and Fig 20 2 The certificate specifies that the indicator should be connected to a limited energy 4 20mA...

Страница 29: ...Fig 21 2 The four vents at the rear of the indicator enclosure shown in Fig 22 must not be obstructed 3 The certificate specifies that the indicator should be connected to a limited energy 4 20mA circ...

Страница 30: ...ne 22 The panel enclosure containing the indicator should be located where the minimum amount of dust will accumulate on them and the installation should be as specified in section 5 of this instructi...

Страница 31: ...X 4 ETL cETL certification for installations in USA and Canada A4 0 cETL Mark For installations in the USA and Canada the BA307NE and BA327NE indicators have ETL and cETL Ex nA and Ex tc approval Cont...

Отзывы: