Appendix
Operating manual AVERO Motion
V2.6 E
Page 40 of 48
13.7.5
Changing the disinfectant hose
Picture 1:
The picture shows the
disinfectant hose as well
as the elbow fitting on the
tub.
Picture 2:
Unscrew the
disinfectant hose from
the elbow fitting.
Picture 3:
The picture shows the
unscrewed disinfectant
hose.
Assembly of the disinfectant hose is done in the reverse order!
13.7.6
Changing the eccentric screw fittings (drainage and
overflow)
Picture 1:
The picture shows
the eccentric
cover.
Picture 2:
To loosen the cover,
pull it in an upward
direction.
Picture 3:
The picture shows
the removed cover.
Picture 4:
The rotary actuation
must be positioned
horizontally (Viega
logo must be
recognizable) before
the drip plug can be
loosened.
Picture 5:
Bring the eccentric
plugs to an upright
position.
Picture 6:
The upward pulling
motion allows the
removal of the
eccentric plugs.
Picture 7:
The picture shows
the removed
eccentric plugs
Warning:
When fitting the drip plug into its casing,
ensure it is correctly locked.
Assembly of the drip plugs is done in the reverse order.
Содержание AVERO Motion
Страница 33: ...Appendix Operating manual AVERO Motion V2 6 E Page 33 of 48 13 3 Wiring Diagram 13 3 1 Connection at the side...
Страница 34: ...Appendix Operating manual AVERO Motion V2 6 E Page 34 of 48 13 3 2 Connection at the rear...
Страница 35: ...Appendix Operating manual AVERO Motion V2 6 E Page 35 of 48 13 3 3 Connection at the floor...
Страница 36: ...Appendix Operating manual AVERO Motion V2 6 E Page 36 of 48 13 4 Declaration of Conformity...
Страница 47: ...Appendix Operating manual AVERO Motion V2 6 E Page 47 of 48 Your notes...
Страница 48: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...