Prism Medical Ltd. V1-2015
41
Guidelines and manufacturer declaration
– electromagnetic resistance
The AVERO Premium Plus is intended for use in the electromagnetic environment specified below.
The customer or user of AVERO Premium Plus should ensure that it is used in such an environment.
Interference
Resistance
IEC 60601-
test level
Consistency
level
Electromagnetic
Environment Guidelines
Performed RF
IEC 61000-4-6
Radiation HF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
10 Vrms
3 V/m
Portable and mobile RF communications
equipment
should be used no closer to AVERO Premium
Plus, including cables, than the recommended
separation distance
calculated according to the applicable
transmission frequency.
Recommended separation distance
d=0.35√P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
Where P is the nominal power of the transmitter in
watts (W) according to to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Determines the field strength of stationary radio
transmitters
by electromagnetic location measurement,
a
–
should be less than the CONCORD level in all
frequency ranges.
b
In the vicinity of equipment marked with the
following
symbol interferences are possible:
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These instructions cannot be applied in all situations under certain circumstances. The propagation of
electromagnetic waves is influenced by absorption and reflection from structures, objects and people.
a
The field strength of stationary transmitters, such as base stations of cordless telephones and land mobile radios,
amateur radio, AM and FM stations as well as radio and TV stations cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic survey site is
recommended. If the measured field strength in the environment in which the AVERO Premium Plus is to be used
exceeds the applicable RF compliance level, particular attention must be paid to be able to ensure the normal
operation of the AVERO Premium Plus. If abnormalities are observed, additional measures may be necessary, such
as reorienting or relocation of AVERO Premium Plus.
b
In the frequency range of 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Содержание 810142501
Страница 1: ...1 Plumbing CSA B45 5 11 IAPMO 7124 2011 Electrical CSA C22 2 No 60601 1 ...
Страница 6: ...Prism Medical Ltd V1 2015 6 2 Introduction WARNING ...
Страница 31: ...Prism Medical Ltd V1 2015 31 11 Troubleshooting and customer service Problem Solution ...
Страница 35: ...Prism Medical Ltd V1 2015 35 ...
Страница 36: ...Prism Medical Ltd V1 2015 36 ...
Страница 37: ...Prism Medical Ltd V1 2015 37 ...
Страница 46: ...BEKA Hospitec Ltd Am Rübenmorgen 3 D 35582 Wetzlar Telephone 49 0 6 41 9 22 20 0 Fax 49 0 6 41 9 22 20 20 ...