Prism Medical Ltd. V1-2015
14
Connection to the power supply network:
When connecting the device to a power supply, the specific CSA standards must be observed.
The connection of the AVERO Premium Plus is connected by means of a distribution box IP X4,
120V AC/60 Hz, 3x1.5 mm
2
, RCD 30 mA, FI-safety switch 30 mA potential equalization according
to DIN VDE 0107.
Please note that the connection must be performed by a qualified electrician.
Ground wire:
When using devices of protection class l, the quality of the protective ground conductor and
maintenance is important. Follow local electrical regulations.
WARNING:
Operation:
The knob regulates the water temperature. If you turn the knob clockwise until the end, the water
runs cold. If you turn the knob counter-clockwise, the water runs warm to hot. If you require a hotter
water temperature, you have to press the safety lock.
Safety lock
Cold
Warm
To avoid the risk of electric shock, this unit must be
connected to a supply network with a ground wire.