User’s Guide
Chapter 2
Document 609909-AA
2-4
Biomek 2000
Side Loader (SL)
Verify that the voltage selector is set for correct voltage by a Beckman Representative
at the time of installation.
•
DO NOT touch, change, or otherwise interfere with labware on the worksurface,
stacks, or in the SL hand unless the system instructs you to do so.
•
The arm can move suddenly. Stay clear of the arm and hand during operation.
•
Keep the area around the SL clear of any obstructions to avoid collisions.
•
There is a Stop button at the base of each Stack Area. If an emergency stop is
required, pressing this on each stack area of your SL button will freeze the SL
instantly.
CAUTION
The Stop button on the SL will not stop the workstation,
nor will the Stop button on the workstation stop the SL.
NOTE
You may lose data and be forced to quit the run in progress if you
press the SL Stop button. Click on the Stop button on the Run win-
dow to stop the instrument if it is not an emergency. Use of the SL
stop button may result in spillage or splattering of liquids in transit.
SL Incubator
•
Operating the incubator at 4° C for longer than twenty-four hours may result in ice
formation within the chamber, adversely affecting temperature equilibrium and
assay results.
•
Do NOT place the incubator in stack location "C" as the incubator may inhibit
access to the Stop button of the SL in the event of an emergency.
•
Allow enough time for the incubator to reach the desired temperature to ensure
reliable assay results (typically less than one hour, depending on the circulator and
desired temperature).
•
Use 50% polyethylene glycol and 50% water as the conductive fluid in the water
circulator.
•
Maximum inlet pressure to the SL Incubator is 5 psig (34.5 kPa).
•
When cleaning the incubator wear protective gloves and eyewear.
Содержание Biomek 2000
Страница 8: ...Notice vi Biomek 2000 ...
Страница 24: ...User s Guide Chapter 1 Document 609909 AA 1 10 Biomek 2000 ...
Страница 58: ...User s Guide Chapter 3 Document 609909 AA 3 28 Biomek 2000 ...
Страница 61: ...Chapter 4 User s Guide Document 609909 AA Biomek 2000 4 3 Select Lab Book ...
Страница 63: ...Chapter 4 User s Guide Document 609909 AA Biomek 2000 4 5 ...
Страница 80: ...User s Guide Chapter 6 Document 609909 AA 6 2 Biomek 2000 ...