15-March-06

Creating and Modifying Vi-Cell Cell Types

When you begin using the instrument, it is best to select the 

Default Cell Type

to analyze your sample.  Then, you can adjust these parameters to fit the

specific cell line you are measuring and create a unique cell type.  These cell
types will show up in everyone’s user account and there should be only 

one cell

type for each cell line

 that we use in the lab.  Please talk with other lab

members to set the parameters for a given cell line so that everyone can use
the same cell type for a given cell line.

For additional details, please see the Vi-Cell User Manual or the cell type power
point tutorial, which is saved on the desktop of the Vi-Cell computer with the
extension pps.

To access the cell types:

1.

 

choose 

File 

from toolbar then click on 

Cell Type

 and you will be given

the option to add, modify, or delete a cell type.

2.

 

choose 

File 

from toolbar then click on 

Image-Reanalyze 

and you will be

given the option to add, modify, or delete a cell type.

3.

 

choose 

Instrument-Reanalyze 

from toolbar then you will be given the

option to add, modify, or delete a cell type.

Sample Prep Options

Parameter

Range

Default

Details

Aspiration

Cycles

1 to 7

3

# of times sample is mixed in

syringe; set according to cell

sensitivity & clumpiness

Trypan Blue

Mixing

1 to 7

3

# of times sample and trypan blue

are mixed in syringe

Image Analysis Parameters

The ranges given below are the normal recommended ranges for each
parameter and default value refers to the parameter value in the 

Default Cell

Type

.  The default values are a good starting place and you should tweak

individual parameters to determine their effects on the output values.

For the cell brightness parameter, you will need to observe the image in binary
mode to see the difference between parameter values.  Right click on the
image, then 

turn off annotation

 and then choose 

Binary Mode.  

Return to

normal view for all other parameters.

Содержание Vi-Cell XR

Страница 1: ...tem administrator In addition everyone will be given a user account and password to log on to the instrument There will be a main system log in which user activity will be recorded So if you allow someone else to use your account then you will be the responsibility party in the use log if something happens to the instrument UROPs will not be given user accounts and will be the responsibility of th...

Страница 2: ...isplayed in the main window Camera Image allows for the viewing of sample images during and after a run Autosampler Queue shows the completed samples along with the list of samples in the queue this is where you can modify remove samples that are already in the queue Cell Types displays windows with the information for each specific cell type this is where you create or remove new cell types Biopr...

Страница 3: ...ue to enter samples so it saves time to start the queue first 5 From the navigation menu choose Autosampler queue to see your samples in the queue From this screen you can also edit remove samples in the queue while a run is in progress 6 Click on Start queue to begin sample analysis Once the run begins your sample will disappear from the queue and you will only see it on the main screen The botto...

Страница 4: ...roll down to click on reanalyze At this point the default cell type window will open up and you can adjust the necessary parameters and click ok The cell images will then be analyzed using the new parameter values and the data will be displayed on the right side of the screen Reanalyzing a sample does not overwrite the original data and the new data is not automatically saved So you need to choose...

Страница 5: ...spending your cells in a solution that is also hazardous you need to write that information on the waste card as well every time you use the cell counter As for the instrument reagents these will be ordered from Beckman Coulter by the lab manager and stocked near the instrument Only the Beckman Coulter reagents are compatible with this instrument Do NOT use any other brand of trypan blue If we are...

Страница 6: ...ed on spot brightness spot area Live Cell Reject Non Circular Objects Separate Cells from Background Based on brightness sharpness Vi CellXR Cell Image Analysis Scheme Not Viable Debris Meets Min Circularity Viable Decluster Cells Dead Cell ...

Страница 7: ...delete a cell type 2 choose File from toolbar then click on Image Reanalyze and you will be given the option to add modify or delete a cell type 3 choose Instrument Reanalyze from toolbar then you will be given the option to add modify or delete a cell type Sample Prep Options Parameter Range Default Details Aspiration Cycles 1 to 7 3 of times sample is mixed in syringe set according to cell sensi...

Страница 8: ...o include more objects Increase to include less objects Spot Brightness 75 Determines cell viability Increase value more cells counted as dead Decrease value more cells counted as live Spot Area 1 20 5 Ratio of the bright spot to total area Determines cell viability Increase value more cells counted as dead Decrease value more cells counted as live Minimum Circularity 0 0 5 0 Excludes dead cells s...

Отзывы: