Professional Development
Supplies History Logs
There are two history logs associated with Supplies. One is located under Event Logs
and the other is located under Data Summary Logs.
Supplies Event Log:
This log records any changes or alerts associated with the
supplies.
Supplies Data Summary Log:
This log records all the information on each new
container set up on the System.
Step
Action
3
Scan or manually enter the barcodes, into
Bar Code
fields 1 and 2, and select
OK
.
4
On the Supply field, choose the reagent to be replaced from the drop-down list.
5
On the displayed message, select
OK
to verify that the reagent has been replaced
or
Select
Cancel
to go back to the previous screen.
6
You would repeat steps 3 to 5 above for each remaining supply.
7
Select
Details
from the local navigation bar to display information for each supply.
Note:
Slidestainer Supplies
Details indicates the supplies required for the active
staining protocol set for the instrument.