![Beckman Coulter UniCel DxH Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/beckman-coulter/unicel-dxh/unicel-dxh_user-manual_2719769029.webp)
PN: B11956AA (March 2012)
Page 29 of 46
Professional Development
Staining Samples Manually (Stain-Only Priority)
Risk of personal injury and contamination. The processing areas could contain prepared smears
and/or broken glass with biohazardous material. Handle with care and avoid skin contact.
Some samples may require manual staining, such as small blood draws (pediatric)
taken with microcollection devices.
Follow these steps to manually stain a sample.
Step
Action
1
Fill in an empty basket with slides ready to stain (painted end on top so as not to
stain the sample information).
2
Place the basket back in the I/O drawer in any open/empty position from 1 to 6.
3
Press the number where the basket is placed.
4
Push the drawer in for staining to begin.