![Beckman Coulter TQ-Prep Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/beckman-coulter/tq-prep/tq-prep_reference_2719765059.webp)
PN 4237395AA
B-4
OPERATOR CERTIFICATION PROCEDURE
PIPET CALIBRATION PROCEDURE
7
Slowly release the plunger to overdraw
the required volume.
8
Remove the tip from the deionized or
distilled water while touching the tip to
an inside wall to remove any adhering
liquid.
9
While touching the tip to the tared
weigh boat, slowly dispense the water
by depressing the plunger to the first
stop only. There must be liquid
remaining inside the tip.