OPERATION PRINCIPLES
TYPES OF PRIMING
3
New Reagent
r
This prime dispenses enough reagent to replace the reagent currently in all the syringes,
in the tubing from the syringes to the dispensing head and in the tubing from the
syringes to the reagent containers.
r
See Chapter 4, Specifications/Characteristics, for amounts of dispensed reagent.
r
Press
to request this prime.
Automatic
This prime dispenses 20 µL of each reagent automatically when you request a function
(sample processing, sample processing with preincubation, Gravimetric Validation, user
requested prime or new reagent prime) after the instrument has been inactive for at least 12
hours.
Содержание TQ-Prep
Страница 4: ...PN 4237395AA iv REVISION STATUS ...
Страница 8: ...PN 4237395AA viii CONTENTS ...
Страница 17: ...PN 4237395AA 1 6 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 43: ...PN 4237395AA 3 4 OPERATION PRINCIPLES TYPES OF PRIMING ...
Страница 47: ...PN 4237395AA 4 4 SPECIFICATIONS CHARACTERISTICS PERFORMANCE CHARACTERISTICS ...
Страница 49: ...PN 4237395AA A 2 LOG SHEETS ...
Страница 51: ...PN 4237395AA A 4 LOG SHEETS ...
Страница 53: ...PN 4237395AA A 6 LOG SHEETS ...
Страница 55: ...PN 4237395AA A 8 LOG SHEETS ...
Страница 69: ...PN 4237395AA B 14 OPERATOR CERTIFICATION PROCEDURE TQ PREP WORKSTATION PRECISION PROCEDURE ...
Страница 71: ...PN 4237395AA B 16 OPERATOR CERTIFICATION PROCEDURE TQ PREP WORKSTATION PRECISION PROCEDURE ...
Страница 73: ...PN 4237395AA 2 REFERENCES ...
Страница 77: ...PN 4237395AA 4 GLOSSARY ...
Страница 85: ......