CHAPTER 3
Operation
Introduction
WARNING
Handle infectious samples according to good laboratory procedures and methods
to prevent spread of disease. Because spills, operator error, or tube failure may
generate aerosols, observe proper safety precautions for aerosol containment.
Do not run toxic, pathogenic, or radioactive materials in this centrifuge without
taking appropriate safety precautions. Biosafe containment should be used when
Risk Group II materials (as identified in the World Health Organization
Laboratory
Biosafety Manual
) are handled; materials of a higher group require more than one
level of protection.
WARNING
The centrifuge must not be used in the vicinity of flammable liquids or vapors,
and such materials should not be run in the centrifuge. Never bring any flammable
substances within the 30-cm (1-ft) boundary surrounding the centrifuge. Do not
lean on the centrifuge or place items on the centrifuge while it is operating. During
operation you should not come within the 7.6-cm (3-in.) clearance envelope
except to adjust centrifuge controls, if necessary.
Run Procedure
CAUTION
The Microfuge 16 centrifuge and FX241.5P and FX121.5P rotors were developed,
manufactured, and tested for safety and reliability as part of a Beckman Coulter
centrifuge/rotor system. Safety or reliability of the centrifuge cannot be assured
if used with any other rotor.