Troubleshooting

Diagnostics screen

The Diagnostics icon is in the Communications Setup sub-menus.
The Diagnostics screen shows information that may be useful in the analysis of instrument failures.
Information shown includes details about instrument calibration, laser current, firmware version,
battery voltage and battery charge status.

Error descriptions

Table 3

 describes the kinds of errors that can occur.

Table 3  Errors

Error type

Description

Stop

A stop error causes all operation to stop until the error is corrected. The current sample is
aborted.

Flow

A flow error causes the counter to stop sampling, and an error message shows in the display.

Hardware failure

Hardware failures are indicated by an error message in the display. Record any fault codes
that show.

Warnings

A warning occurs when a subsystem does not function correctly. When a warning occurs, the
instrument continues to operate and a warning level indicator shows in the display. When the
warning condition is cleared, the indicator goes out of view.

Table 4

 describes the kinds of warnings that can occur.

Table 4  Warnings

Warning

Description

Laser current

Occurs when the laser current exceeds ±30% of the nominal level.

Buffer full

The instrument is powered on and the buffer is full. Old data will be overwritten.

Flow system

A Flow Alarm warning occurs when the firmware is unable to maintain a steady state.

Overconcentration Occurs when concentration limits of the sensor are detected. The warning clears when

concentration values return to normal.

Calibration failure

Occurs when the calibration signal registers a sensitivity change of ± 10% size error in the
first channel.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Description

Item no.

Adapter for zero count filter

2089398 

Charge cradle

2089380-01 

Power supply

230-300-1000 

USB cable

460-400-0002 

Zero count filter

VP212808 

Carrying case

2089328-01 

18

   

English

Содержание HHPC2+

Страница 1: ...DOC026 97 80271 HHPC 6 HHPC 3 HHPC 2 10 2013 Edition 7 User Manual Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...

Страница 2: ...s in a minute USB support Simple data export to memory stick or direct connect to PC via USB cable No software required Ethernet support Access particle data over network with an internet browser such...

Страница 3: ...out limitation direct incidental and consequential damages and disclaims such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application...

Страница 4: ...ions 1 The equipment may not cause harmful interference 2 The equipment must accept any interference received including interference that may cause undesired operation Changes or modifications to this...

Страница 5: ...quality in Clean rooms Manufacturing processes Pharmaceutical production Three models of the instrument are available The main differences are given in Table 1 The cradle keeps the instrument in an u...

Страница 6: ...radle accessory optional 4 AC adapter Data and power connections Figure 2 shows the locations of the data and power connections The ethernet port is available only on the cradle For more information o...

Страница 7: ...h drive or memory device 9 Ethernet port 5 Threaded tripod mount User interface and navigation User interface The LCD display screen and the 7 button keypad on the front of the instrument operate as t...

Страница 8: ...e for some menus Information on the help screens helps the user set up and use the instrument To show the information on a help screen highlight the icon in the lower right corner of the display then...

Страница 9: ...ption 1 Push the Menu button 2 Push the right or left arrow buttons to highlight an icon Push the Up or Down arrow buttons to move through sub menu options 3 Push the Select button to accept the highl...

Страница 10: ...sed vehicle in direct sunlight Failure to obey this precaution can cause the battery to overheat and cause a fire or explosion Note Potential instrument damage Do not use the instrument in locations w...

Страница 11: ...is put in the user has only Operator level access If the Security feature is set to off the box is not checked a user has both Operator and Administrator level access Change Password The default pass...

Страница 12: ...unt process until there are no new counts 4 Remove the zero count filter for normal operation Sample setup Refer to the help screen for more information Change the settings for how the instrument runs...

Страница 13: ...cess The instrument begins the sample process with the default or stored Sample Setup values While the instrument takes samples the display shows the sample status current sample number and the sample...

Страница 14: ...sh the up or down arrows to scroll through the data The data scroll in sequence from the current record The data include the date and time of collection the record number currently in view channels an...

Страница 15: ...he IP address Subnet Mask and Gateway address To do this do one of the following a Put the IP Address Subnet and Gateway data in the appropriate fields OR b Select the DHCP check box If the DHCP box i...

Страница 16: ...C adapter plug to the AC adapter connector of the optional cradle then attach the instrument to the cradle Connect the AC adapter plug to the AC adapter connector on the bottom of the instrument An am...

Страница 17: ...l indicator shows in the display When the warning condition is cleared the indicator goes out of view Table 4 describes the kinds of warnings that can occur Table 4 Warnings Warning Description Laser...

Отзывы: