Operation
3-14
6. Check that all parameters are correct and the door is shut and latched,
then press [
ENTER
], then [
START
].
WARNING
!
Never attempt to override the door interlock sys-
tem while the rotor is spinning.
!
CAUTION
Do not lift or move the centrifuge while the rotor is
spinning
7. Wait for the set time to count down to zero, or end the run by pressing
either the [
STOP
] key or the [
FAST STOP
] key.
8. After the rotor stops spinning and the [
OPEN DOOR
] light comes on,
press the [
OPEN DOOR
] key to release the door latch; open the door.
9. Unload the rotor according to instructions in the applicable rotor
manual.
!
CAUTION
If disassembly reveals evidence of leakage, you
should assume that some fluid escaped the rotor.
Appropriate decontamination procedures should be
applied to the centrifuge and accessories.