5. Water distiller with built in tank automatic 

controller principle of operation

5.1.  The water distiller is connected to the mains using a plug (for 1.5 kW version of BEAR 

B 1104 and BEAR B 1204) or an input circuit breaker (BA) (for 3 kW version of BEAR B 
1104 and BEAR B 1204 and all versions of BEAR B 1110, BEAR B 1210, BEAR B 1125) , 
which should be installed by the Customer and is intended for the overload / short cir-
cuit protection of electric circuits; it is selected by the 10xIn overcurrent tripping setting 
(where In is a rated current).

5.2.  The water distiller is started by setting the toggle POWER switch to ON, at the same 

time the POWER LED lights and supply voltage is applied to the automatic device that 
activates raw water supply valve  21 (Figure 1), at the same time the INLET WATER led 
lights. 

5.3.  When the water level in the evaporator 2 (Figure 1) reaches the bottom sensor 9, at the 

same time the NORMAL WATER LEVEL LED lights, the automatic device activates the 
heater (s) 17 (Figure 1), at the same time the switch HEATER LED lights. 

5.4.  When the evaporator water level lowers below the admissible level (below the level 

sensor 9) the automatic device switches the heater OFF, at the same time the switch 
led HEATER turns off. 

5.5.  The Low water thermostat sensor 10 (Figure 1) ensures an additional emergency shut-

down of the water still in the case of water level lowering and the heater temperature 
rise over 150 ̊С. 

5.6.  When the distillate collector is filled up to the level sensor 11 (Figure 1), the automatic 

device stops the heater 17 (Figure 1) and shuts raw water supply valve 11 (Figure 1) off, 
at the same time HEATER LED and INLET WATER LED turns off.  The FULL RESERVOIR 
LED lights. 

5.7.  When distillate is drained from the collector, The FULL RESERVOIR LED turns off, the 

automatic controller activates raw water supply valve 21 and the heaters 17 (Figure 1) at 
the same time the INLET WATER and HEATERS LEDS lights

5.8.  When water overflows in the evaporator 2 (Figure 1) or equalizer 7, the water level 

reaches the top-level control sensor in the evaporator 7 or level sensor in the equaliz-
er 24, the automatic controller switches the heater 17 and raw water supply valve 21 
OFF and the HIGH WATER LEVEL LED will be lit. To continue the operation, you should 
switch toggle POWER switch OFF, eliminate the cause of overflow and switch the tog-
gle POWER switch to ON. 

5.9.  Upon completion of operation you should cut the water distiller of the mains. Switch 

toggle POWER switch and input circuit breaker off.

.

Содержание A 1104

Страница 1: ...Water Distillers OPERTION MANUAL A 1104 A 1110 A 1125 A 1204 A 1210...

Страница 2: ...ke modifications to the design of the water distiller Be careful water distiller surfaces and distillate can be extremely hot while operating Use only original accessories and parts heaters valves con...

Страница 3: ...controlled by personnel all the time while operating Make sure that distillate receptacle is big enough and not full before starting operation 2 General information 2 1 Water distillers BEAR B 1104 BE...

Страница 4: ...Due to heaviness of BEAR B 1110 and BEAR B 1125 those models should be unpacked by 2 persons For BEAR B 1110 and BEAR B 1125 the easiest way is to disassemble the wooden crate and then remove the pal...

Страница 5: ...n hose inlet hose length 2 meters 1 Water connection hose drain hose length 1 5 meters 16x25 mm 2 Fitting with male thread 12 mm 2 Silicone tube for constant distillate withdrawal 15x3 mm length 1 met...

Страница 6: ...1 Fitting with male thread 12 mm 1 Metal water connection hose inlet hose length 1 5 meters 1 Water connection hose drain hose length 1 5 meters 2 Silicone tube for constant distillate withdrawal 16x2...

Страница 7: ...ler design the structure component view is schematic not taking into considera tion features of the water stills Figure 2 shows the fitting grounding terminal connection diagram 4 Water distiller with...

Страница 8: ...changer inlet 16 Evaporator inlet 17 Heater 18 Equalizer drain pipe 19 Heat exchanger 20 Condenser drain pipe 21 Raw water supply valve 22 Collector distillate level pipe 23 Evaporator top level senso...

Страница 9: ...vaporator 2 steam forms then separated and enters in the condenser 6 via the channel 5 The separator consists of the steam floe splitter 3 and the nozzle 4 located inside the channel 5 4 4 Steam conde...

Страница 10: ...utomatic device switches the heater OFF at the same time the switch led HEATER turns off 5 5 The Low water thermostat sensor 10 Figure 1 ensures an additional emergency shut down of the water still in...

Страница 11: ...sideration features of the water stills 6 Water distiller without built in tank design and principle of operation 1 Housing 2 POWER switch 3 Electric heater s 4 Low water thermostatic sensor 5 Distill...

Страница 12: ...ld be used to collect distillate 7 Before installation Installation of the equipment must be done only by qualified personnel and only after reading this operating manual 7 1 The water distillers are...

Страница 13: ...IGH WATER LEVEL red LED will light up 8 6 Install gaskets in the filling hose nuts if they ar not pre installed at the factory 8 7 Connect the filling hose to the fitting 25 Figure 2 8 8 Connect the f...

Страница 14: ...14 Connect the water still grounding terminal to the PG bus with flexible copper wire at least 6 mm2 in cross section Connect the external circuits to the power supply network in accordance with Figur...

Страница 15: ...re Equalizer is about 70 80 degrees Celsius 9 5 Install water distiller on the table using the level gauge and avoiding slants 9 6 Attach the silicone tube L 0 5 m to the distillate drain pipe 5 9 7 C...

Страница 16: ...s an overflow of water and turn off the water still and the HIGH WATER LEVEL LED will be lit 10 7 To start the water distiller push the POWER button located on the front panel and this button should b...

Страница 17: ...d down and then push the lock button located on the thermal switch housing Pos 13 Figure 5 11 6 To start the water still push the POWER button located on the front panel and this but ton should be lit...

Страница 18: ...splitter 3 Figure 1 12 6 6 Remove the steam flow splitter 3 12 6 7 Flush with water the walls and bottom of the evaporator to remove salt con taining sediments You can use the filling hose for supplyi...

Страница 19: ...12 6 28 Make sure that the sensors 9 23 Figure 1 is rigidly fixed and do not contact the walls 12 6 29 Assembly the water still performing steps 4 3 in reverse order 12 6 30 The water still can be use...

Страница 20: ...evaporator with water add cleaning solution for descaling Close the raw water supply valve Bring to a boil and boil for 10 20 minutes Then after 60 minutes drain the solution 13 5 Rinse with water an...

Страница 21: ...ARS Guarantee mentioned above is limited for the heaters for 6 months 16 2 Service Maintenance of the distillers should be performed by qualified personnel Distiller serial number_____________________...

Страница 22: ...Contact Info Location 5115 Douglas Fir Road Suite G Calabasas California 91302 USA Phone 424 259 0816 E mail info bearinnovation com...

Отзывы: