U
SER
M
ANUAL
F
LOW
A
NALYZER
TM
Page 32 / 84
7.9 Menu Display
The following parameters can be viewed and changed in the
menu display:
Calibrations
Gas type and standard
Trigger
Language
HW Activation
System information
7.9.1 Specification of the
Menu Display
(1)
Calibrations.
The oxygen sensor as well as all pressure
and flow sensors and the
MultiGasAnalyzer
TM
MGA-
3050
can be calibrated from this submenu. The offset
calibration for pressure and flow can also be started by
pressing the
Zero!
button.
(2)
Gas type/Standard
. The gas type and standard can be
specified in this submenu.
(3)
The settings in the
Trigger
submenu are used to
measure the respiratory values. By choosing different
respiration modes standard triggers can be selected.
(4)
By selecting a
Filter
the displayed values of the Screen
can be averaged over a certain time.
(5)
Language
. Select the desired language here.
(6)
In the
HW Activation
submenu you can see if the USB
port or the communication to the MGA-3050 multi gas
sensor is enabled.
If the
FlowLab
TM
software or the integration of the gas
sensor was ordered at a later date, you will need to enter
a clearance code before a connection can be made.
(7)
System info
contains information on the software and
hardware versions, as well as data of the last factory
calibration.
Pressing the keys 2 and 3 simultaneously will hide all
menu content that impacts the measurements. This
avoids unwanted changes of the settings.
1
2
3
5
6
11
12
7
4
10
9
8