MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 38 of 144
The available AZT window values are;
Disable
± 0,5e
± 1e
± 3e
This parameter must be disabled or carefully programmed at the silo weighing, tank
weighing and automatic weighing applications for preventing unwanted zeroing.
Default is ‘Disable’.
Tare
If this function is enabled, the weight is tared when the tare command is received.
Additional conditions for taring are positive gross weight and no motion. Multi-taring is
possible at MX08 AD instrument.
(Not available in Count Mode)
Default is ‘Enable’.
Save Tare at Power Off
If this function is enabled, tare value is stored at power off and instrument starts up in Net
mode at power on.
(Not available in Count Mode)
Default is ‘Enable’.
5.3.2.
Calibration
Scale build, operation mode selection and scale calibration is performed in the calibration
tab of xFace as seen in Figure 4.3. Please follow the procedure described in the following
sections in sequence.
This block allows the user to select operation mode and polarity.
Write to Digitizer:
Click this button to save the operation mode and polarity to selected
Digitizer.
Scale setup
Calibration block:
This block allows the user to make calibration with test weights.