MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 37 of 144
Motion Detection
This parameter defines the motion detection window which determines the weighing is
stable or dynamic. If the stable weight status is not required, this parameter can be
disabled.
The available motion detection window values are (Weighing / Count);
Weighing and Force Mode:
± 0.3e
± 0.5e
± 1e
± 2e
No motion detection
Default is ‘± 0.5e’.
Count Mode:
± 60
± 100
± 200
± 400
No motion detection
Default is ‘± 100’.
Power On Zero
This parameter enables automatic power on zeroing at the power on the instrument. This
automatic zeroing is done if the total zero drift from the calibrated zero signal is in the
selected range of the capacity.
The available power on zeroing windows in the percentage of the scale capacity are;
Disable
± 2%
± 10%
This parameter must be disabled or carefully programmed at the silo weighing, tank
weighing or automatic weighing applications for preventing unwanted zeroing at power on.
If the weight is not in zeroing range, instrument is powered on without zeroing.
(Not available in Count Mode)
Default is ‘Disable’.
Zeroing Range
Zeroing of the scale is performed if the difference between load cell signal value and
unloaded load cell signal value at the calibration is in the selected percentage of the scale
capacity. Zeroing can be done by zeroing command when the scale is stable.
The available zeroing ranges in the percentage of the scale capacity are;
Disable
± 2%
± 20%
± 40%
Default is ‘± 2%’.
Auto Zero Tracking
AZT automatically re-adjusts the scale to zero for compensating defined small deviations
around the center of zero. AZT is uses the range of zeroing and is not performed if this
range is exceeded.
(Not available in Count Mode)