5.9.3
WINDOW P1
Window point 1 (for the WINDOW mode) is selected in mm using the arrow keys. The point must lie within the
measuring range (greater than start of measuring range Sdc
+2x hysteresis
).
5.9.4
WINDOW P2
Window point 2 (for the WINDOW mode) is selected in mm using the arrow keys. The point must lie within the
measuring range (less than end of measuring range Sde
1
-2x hysteresis
).
5.9.5
LEVEL
The output level can be inverted with
active high
or
active low
here.
1
According to data sheet chapter 6.1
NOTE
If MEAS TYPE is changed, the switching output settings are discarded. The
factory settings are restored for DIGITAL OUT = End of measuring range Sde