GF Health Products, Inc.  - www.grahamfield.com

Zenith Series Service Manual 999-0831-190C DEC. 2013 

19

ZENITH SERIES

ASSIST DEVICES

Model APS98174, APS981741,

and APS981742

SECTION B

Содержание Slide-W-I-D-E Zenith 9000APS

Страница 1: ...l sections pertaining to the bed model before use GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 ETL UL CSA APPROVED UL 60601 1 IEC 60601 2 38 CSA 22 2...

Страница 2: ...kaging and warranties contained in this Service Manual are based on the latest product information available at the time of printing The most current product information can be found online at www gra...

Страница 3: ...delenberg Wallsaver Brackets 13 Plugging in Your Staff Nurse Control 14 Pendant Hand Controller Operation 15 Staff Nurse Control Operation 16 Bed Operations Chair Position 17 Bed Operations Trendelenb...

Страница 4: ...g blankets pads wiring etc Beds manufactured by BasicAmerican Medical Products are designed for use within an institutional healthcare environment i e assisted living skilled nursing tran sitional car...

Страница 5: ...bed components acces sories that correspond with the needs of the patient resident and the particular bed you have purchased Matching the correct bed component that correlates with the regulatory gui...

Страница 6: ...ages properly X Check the bottom foot pads on the foot end casters making sure they are clean of debris and undamaged Replace if needed X Check all casters to ensure that they roll properly and are un...

Страница 7: ...TINE A L ADMINISTRATION D OXYGENE DE TYPE NASAL OU MASQUE OU AVEC UNE TENTE RECOUVRANT SEULEMENT LA MOTTIE AVENT TETE DU LIT LES COTES DE LAS TENTE OXYGENE NE DOIVENT PAS SE PROLONGER PLUS DAS QUE LA...

Страница 8: ...GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 ZENITH SERIES ZENITH APS 9000 Bed Set up Instructions SECTION A 8...

Страница 9: ...ect the Power Supply Cord for any cuts and or damage Check to see all actuator motor cables are routed and connected properly to the control box DISCARD 1 Middle Foam Block 2 Large Block 3 Large Cable...

Страница 10: ...nsider this the outside of the board Position 2 mounting tubes on the inside of the footboard with L facing outward Align the top hole of the mounting tubes with the top holes in the footboard Insert...

Страница 11: ...76 OR 80 CONFIGURATIONS SEE PAGES 10 AND 12 1 Determine if you need to position your bed in an 80 or 76 configuration 80 outside holes 76 inside holes see Closeups A B 2 On the foot end carefully squ...

Страница 12: ...MOVAL 1 To remove or move the Wireform Wallsaver to a new position raise the wallsaver off the floor until the end tabs are horizontal 2 Squeeze the ends toward the center of the wallsaver until the e...

Страница 13: ...To remove or move the Wireform Wallsaver to a new position raise the wallsaver off the floor until the end tabs are horizontal 2 Squeeze the ends toward the center of the wallsaver until the end tabs...

Страница 14: ...rol cable through the large round hole c With the text right side up insert the Staff Control Assembly into the cutout slot on the footboard and attach to the board using the two outside small holes a...

Страница 15: ...ent hook A Pendant holster also comes standard with the APS bed BACKLIT PENDANT Linak Model HB8545V3203 61 BAMP Part 999 0806 301 Backlit and includes Underbed Light Button BACKLIT PENDANT 1 HEAD DECK...

Страница 16: ...Knee Chair and HI LO functions press the appropriate top icon 2 3 4 10 or 12 button and the Key button at the same time An orange LED lock symbol will appear under the related icon To Unlock any of t...

Страница 17: ...ey icon 1 and Chair icon 2 or All Locked button 3 at the same time 2 To move the bed to the Chair position foot knee decks angled up head deck angled up and head of bed tilted upward press on the chai...

Страница 18: ...2 at the same time 2 To move the bed to the Trendelenberg position foot end up press on button 3 3 To move the APS bed into the Reverse Trendelenberg position head end up press on button 4 1 2 3 4 NO...

Страница 19: ...GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 19 ZENITH SERIES ASSIST DEVICES Model APS98174 APS981741 and APS981742 SECTION B...

Страница 20: ...ce 2 To set the assist bar back to its upright locked position grab the top of the assist bar with one hand and pivot the assembly upward until the black knob mechanism snaps into place locking the as...

Страница 21: ...AIL FOUR LANYARD ASSEMBLIES AND 2 PHILLIPS SCREWS OR 2 OPTIONAL HEX HEAD SCREWS ORDER BAG 999 0842 905 FOR JUST THE REPLACEMENT HARDWARE THE PLASTIC RAIL CAN BE POSITIONED ON EITHER THE HEAD OR FOOT D...

Страница 22: ...bag Do not remove the 4 four 1 4 20 inserts 3 from the board They will be used again to mount the board to the trapeze support 3 Assemble the Trapeze Support 1 to the main frame tubes using the two C...

Страница 23: ...GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 23 ZENITH SERIES Mechanical Electrical Information SECTION C...

Страница 24: ...d without assist devices or boards 204 lbs Bed height calculated from floor surface to top of sleep deck NOTE All dimensions are in a range of 25 inches GF HEALTH PRODUCTS INC DUTY CYCLE 2 min Continu...

Страница 25: ...ed Control Junction Box Part 999 0831 300 Quantity 1 Power Cable plugs into side port part listed at bottom Head Motor Actuator Part 999 0822 052 Quantity 1 Foot Motor Actuator Part 999 0822 053 Quant...

Страница 26: ...Cable Ground Wire T Cable HB HI LO Ft Port 4 Foot Port 3 HI LO Hd Port 2 Head Port 1 Power Cable Battery Cable Side of Control Box Portable Battery Cable Extension POWER CABLE T CABLE HEAD CABLE FOOT...

Страница 27: ...OOT CABLE HI LO CABLE FOOT HI LO CABLE HEAD BATTERY BACKUP CABLE GROUND WIRE UNDERBED LIGHT CABLE UBL u v w x y z u v v v v w y x z NOTE Cabling for Beds with Underbed Light feature is identical to St...

Страница 28: ...ABLE HEAD BATTERY BACKUP CABLE GROUND WIRE UNDERBED LIGHT CABLE UBL u v w x y z u v v w y x z NOTE Cabling for Beds with Underbed Light and Mounted Battery Backup features are identical to Standard AP...

Страница 29: ...GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 29 ZENITH SERIES SERVICE REPLACEMENT PARTS SECTION D...

Страница 30: ...ttached Hardware Pack 1 Motor 999 0822 052SP OR Head Motor Cable attached Hardware Pack 1 Motor 999 0822 053SP OR Foot Motor Cable attached Hardware Pack 1 Motor Motor Hardware Pack INCLUDES 1 Pkg Sep...

Страница 31: ...with Bezel Cables 1 Unit ZL877000 Staff Control Service Pack INCLUDES 1 Pkg PLEASE SEE PAGE 14 FOR INSTALLATION INSTRUCTIONS 1 Switch Pad with Bezel Cables 1 Shroud Cable Cover 6 6 Phillips Head Trus...

Страница 32: ...ludes control box detachable power cord control cable ground wire screw cable ties installation instruction sheet 5 TO ORDER A REPLACEMENT DOUBLE T CABLE USE SERVICE PACK 999 0806 200SP Includes T Cab...

Страница 33: ...cable ties to re secure to tie rod and mounting brackets 2 1 If the APS bed came with a mounted Battery Backup Pack Part 999 0831 003 you must first unplug the power cord from the wall unit then unplu...

Страница 34: ...GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 34 ZENITH SERIES FRAMING SERVICE PARTS For All Zenith APS 9000 Bed Models SECTION E...

Страница 35: ...ounting hardware installation instruction sheet 3 TO ORDER A REPLACEMENT FOOT DECK USE SERVICE PACK 999 0844 930SP includes all deck mounting hardware installation instruction sheet Re use existing ed...

Страница 36: ...FRAME CAPPING CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON THE BED SEE BOTTOM OF PAGE 31 FOR ORDER NUMBERS HEAD END FOOT EN...

Страница 37: ...ERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON THE BED SEE BOTTOM OF PAGE 31 FOR ORDER NUMBERS CAPPING ON ARMS LEGS 2X 100 4700 018 Rectangular End Ca...

Страница 38: ...GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0831 190C DEC 2013 38 ZENITH SERIES TROUBLESHOOTING All Zenith APS 9000 Models SECTION E...

Страница 39: ...frayed or exposed If it is OK you should not have to replace POWER CORD IS DAMAGED Cut cable ties and immediately replace Power Cord 1 STAFF CONTROL LIGHTS ARE BLINKING NOTHING ON STAFF CONTROL WORKS...

Страница 40: ...lug the Pendant Hand Controller plug directly into the HB port on the Control Box making sure it is fully seated Plug in the Power Cord and test the Pendant ALL BUTTONS WORK Replace the T Cable See pa...

Страница 41: ...If you press the Up or Down HI LO button and you hear a click or buzz but nothing happens Check the HI LO Motors by touching them pushing the HI LO buttons YOU FEEL NOTHING possible bad motor go to th...

Страница 42: ...ips Screws 2 from the inside of the footboard that attach the Staff Control Assembly bevel to the board Set aside screws for later re assembly e Remove the OLD Staff Control Assembly and feed the old...

Страница 43: ...ase price Any repair work or replacement shall not exceed the warranty duration Basic American Medical Products will at its option repair or replace defective product or parts provided that the replac...

Страница 44: ...Lac WI 54937 Customer Service 1 800 365 2338 Fax 920 929 8213 www grahamfield com 2013 GF Health Products Inc All Rights Reserved Basic American Medical Products and Zenith are registered trademarks...