LIBERTY QUICK REFERENCE -
ELECTRONIC TROUBLESHOOTING
The Liberty beds use a 3-motor system with ILCON electronics (after 9/1/2010).
Staff Control is not offered on the Liberty Bed.
GF Health Products, Inc. - www.grahamfield.com Liberty Service Manual 999-0822-190F AUG. 2014
31
LIBERTY SERIES
NO POWER
1. Check Outlet.
2. Unplug Power Cord and Check the following . . .
a. Check Power Cord for Damage or Pinching.
b. Check Short Black Control Box Cable for Damage or Pinching.
c. If any Cables Damaged - Replace Immediately.
PENDANT NOT WORKING
1. Check All Pendant and Cable Connections and retest.
2. Unplug Power Cord.
3. Unplug the Pendant from the Adaptor Cable.
Use a new adaptor cable you know works from another bed. Plug
in the power cord and test:
a. If Pendant works - Replace Adaptor Cable.
b. If Pendant Does Not Work - Unplug the Pendant from Control Box.
Use a new Pendant you know is working from another bed. Plug in
the power cord and test:
i. If Pendant works - Replace old pendant.
ii. If Pendant Does Not Work - Replace Control Box.